                                              Cleaning  &  Disinfection                                                                                            

6. A final wipe down by a clean wipe is necessary to avoid of water 

stains.

Disinfection process

:

1. To disinfect the device when completed the cleaning process.
2. Wipe the probe with a cloth dipped in a disinfectant solution such 

as 70-75% isopropyl (rubbing) alcohol solution: contact time ≥10 

minutes for disinfection.

3. The disinfectant free rinse without residues.

Drying process

1. After cleaning and disinfection process above.
Put the device to air-dry in the room temperature 68-86°F, exposure 
time around 30-60 minutes.

                                                        Care  and  Storage                                                                                                                               

1. Do not drop or crush the device. It is not shock proof.
2. Do not dismantle or modify the device other than changing the 

batteries.

3. Do not store the device in an extreme environment with direct 

sunlight, high/low temperatures or in a location with high humidity 
or severe dust.

Please contact your local agents or distributors if the device does not 

function properly as specified under 

Product Specification

.

Product Specification

                                                  Product  Specification                                                                                                                       

Measuring site

Ear canal

 

Measuring times

3 seconds

Measuring range

90°F ~ 110°F(32.2°C~43.3°C)

Accuracy

96.8°F~102.2°F (36°C~39°C):±0.4°F (0.2°C)

<96.8°F (36°C) : 0.5°F(0.3°C)

>102.2°F (39°C) : 0.5°F(0.3°C)

Resolution

0.1°F

Memory sets

9 sets

Operating condition 60.8°F~104.0°F(16°C~40°C), R.H. of 

15~95% non-condensing

Storage condition

-4°F~120°F(-20°C~50°C), R.H. of 15~95% 

non-condensing

Battery

AAA battery x 2pcs

Battery life

More than 1000 measurements

Auto shut off

After 60 seconds of inactivity

Dimension

102.5mm x 26mm x 26mm

Weight

48g including battery

Warranty

1 year

Guarantee of 

Quality

ISO 13485

ASTM E1965-98

 

Summary of Contents for KI-8130

Page 1: ...planation of Symbols 4 Battery Installation 5 Setting 5 Memory Function 6 Use 7 Cleaning Disinfection 8 Care and Storage 9 Product Specification 10 Trouble Shooting 11 FCC Statement 11 Limited Warranty 12 Lot Number information 12 Table of Contents INSTRUCTION MANUAL Distributed by 180 INNOVATIONS LAKEWOOD CO 80215 Manufactured by K jump Health Co Ltd Made in China www 180innovations com ...

Page 2: ...frared from human body to detect your temperature It can be used with adult or pediatric patients and it can be used in household Safety Instruction WARNINGS 1 Read the entire instruction manual before trying to operate the device 2 The device is delivered non sterile please clean the device before measuring 3 Stop using the device if malfunctions appear or if it operates abnormally 4 Children sho...

Page 3: ...icated 1 Place the probe in scarred tissue or tissue compromised by any disorders of measuring site 2 Patients in trauma or treat with certain drug therapies of measuring site Device Overview Display Overview Quick Start 1 Install batteries into the device Make sure the polarity and of the battery matches as shown in the battery compartment 2 Press and release the power button you will hear a shor...

Page 4: ...e 1 To stop deleting memories do not press the power button when you see dEL 2 Then the device will return to memory mode after 2 seconds then it is ready for measurement Delete memory Power Off Please read thoroughly before using Warning and Caution YYMM Lot Number composed by year Y and month M of manufacture Battery Installation WARNING Use only same type of alkaline batteries to avoid incompat...

Page 5: ...n accordance with the temperature shown as below table Temperature T Backlight T 98 9 Green Backlight 99 T 99 4 Yellow Backlight 99 5 T Red Backlight Cleaning Disinfection 1 To clean and disinfect the device use a cotton swab or pad moistened with 70 alcohol to wipe the Sensor and Probe 2 Use a cloth dampened with water or neutral detergent to clean and dry the device 3 Do not use thinner benzene ...

Page 6: ... the device in an extreme environment with direct sunlight high low temperatures or in a location with high humidity or severe dust Please contact your local agents or distributors if the device does not function properly as specified under Product Specification Product Specification Product Specification Measuring site Ear canal Measuring times 3 seconds Measuring range 90 F 110 F 32 2 C 43 3 C A...

Page 7: ...g measures Reorient or relocate the receiving antenna Increase the separation between the product and the receiver Connect the product into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help REV 20171206 PN 828130XXXX Changes or modifications not expressly approved by the party responsible for compliance c...

Reviews: