Guidance and manufacturer´s declaration – electromagnetic immunity – for EQUIPMENT and SYSTEM that are not
LIFE-SUPPORTING
Guidance and manufacturer´s declaration – electromagnetic immunity
The Infrared Thermometer is intended for use in the electromagnetic environment specified
below. The customer or the user of the Infrared Thermometer should assure that it is used in such
an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
IEC 60601
test level
±6 kV
contact
±8 kV air
±6 kV
contact
±8 kV air
Compliance
level
Electromagnetic environment guidance
Floors should be wood, concrete, or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Guidance and manufacturer´s declaration – electromagnetic immunity
The Infrared Thermometer is intended for use in the electromagnetic environment specified on
the following page. The customer or the user of this Infrared Thermometer should assure that it is
used in such an environment.
Power frequency
(50/60 Hz)
magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial or hospital environment.