25
CONTROL SOLUTION TESTING
STEP 6
Beeps when test is fi nished. The test
result of control solution is displayed on
screen.
STEP 7
Compare the test result with control
solution range which is printed on test
strip container. If the test result is not
within the range, the
2in1.® BASIC
and
strip may worked improperly. Repeat the
control solution test.
STEP 8
Roll the eject wheel to discard the used strip. Test result is
stored in memory automatically.
Control Test
Range 80-120
mg/dL
•
when you are not comply the instructions on this USER’S
GUIDE,
•
expired, contaminated or watered-down control solution,
•
expired or damaged test strip,
•
control solution test is done out of temperature range 20°C
to 25°C,
•
or a problem on the
2in1.® BASIC
Out of range results may be caused by:
If your control solution test results are out of the range
continuously, do not use
2in1.® BASIC
, the test strips and/or
the control solution. Please contact the local distributor.