Disclaimer

 

The infrared temperature measuring device is the human body surface temperature 
screening instrument. The test results only re

fl

ect the real-time temperature evaluation 

of objects or people in a speci

fi

c environment. The selling company and the manufacturer 

shall not bear any responsibility for the direct or indirect loss caused by the use of the test 
results.

 

Installation precaution

 

1. It is recommended to install it indoor free from wind. The cold wind might in

fl

uence the

accuracy in windy outdoor. In this case, we have to increase the environment temperature
compensation, 5° C per time.

2. The temperature measurement area should be protected from sunlight to avoid high results.
3. When it is used outdoor under special circumstances, build a shed on site. The height of

the shed should be more than 2.4m (7.87ft) to ensure the stability and free from wind.
People walk slowly through it to measure the temperature.

Temperature measurement module

 

Temperature sensor:

 

Real time infrared thermo sensor to detect temperature, 
ready in 30 seconds after power on.

 

Distance sensor:

 

The distance sensor starts working when a person or 

object is in 20cm (7.87” in) to it. If the temperature of person 
or object detected is 

≥ 

35.5 

, the voice "Normal 

temperature" is trigged; if it's lower than 35.5 

, or higher 

than the set value with alarm, the voice "Please check 
again" is trigged.

 

When the ambient temperature is lower than 15 

, if the 

temperature of person or object detected is 

≥ 

34.5 

, the voice 

"Normal temperature" is trigged; if it's lower than 34.5 

, or higher 

than the set value with alarm, the voice "Please check again" is 
trigged.

 

Special attention: the distance sensor shall not be blocked, 
otherwise it will repeat the voice broadcast.

 

Power Light:

 

The indicator light is on when the temperature measurement 
module is working. When measuring the forehead temperature, 
the measured result is the most accurate when the middle of 
human eyes is right facing the temperature indicator.

 

Temperature Measurement

 

Power Light

 

Power Light

 

Distance sensor 

Temperature sensor

 

Sketch Map

 

Summary of Contents for 2MTHWT-HMD

Page 1: ...WALKTHROUGH METAL DETECTOR Body Temperature Measurement 2MTHWT HMD Quick Guide Manual Version V6 1 REV06122020...

Page 2: ......

Page 3: ...asure thetemperature Temperature measurement module Temperature sensor Real time infrared thermo sensor to detect temperature ready in 30 seconds after power on Distance sensor The distance sensor sta...

Page 4: ...nd metal detected Shifted by button Power Body temperature and Metal safety pass count Shifted by button Power Temperature sensor Power Select Adjust Confirm A B C D Connection Connect the wires with...

Page 5: ...es press D to back Step 4 press D button then press B to shift and C to adjust to enter the password 1319 press D button to enter below Ambient temperature compensation settings Remark 1 The default e...

Page 6: ...ace When the device is not working properly please restore factory settings Low temperature prompt setting Press D button display CCCC in the main interface press B button to enter the password input...

Page 7: ...mes close within 20cm 7 87 in it triggers the thermal sensor to work When the distance is set to 0100 the temperature result is the most accurate The number of 8190 or 8191 showing in the first line i...

Page 8: ...08 of the first line is the setting to detect small size metals The number 1 of the second line represents to detect small size metals and 50 means not detect small size metals D001 of the first line...

Page 9: ...cannot be measured when the machine is started Treatment wait for 30 seconds after power on and then measure the temperature 3 Low temperature measurement Processing method check whether the value set...

Page 10: ...Body Temperature Measurement Quick Guide V 6 1 1 1 Infrared Temperature Sensor Schematic diagram of temperature measurement...

Page 11: ......

Page 12: ...802 Greenview Dr Suite 200 Grand Prairie TX 75050 USA Toll Free 866 708 5401 Fax 214 988 2858 support 2mtechnology net...

Reviews: