6.4 Determination of acceleration due to 
gravity 

The periods of oscillation 

T

1

 and. 

T

2

 are identical 

at the two points where the curves intersect and 
at these points they are equal to the period of os-
cillation 

T

0

  of  the  calibrated  pendulum,  i.e. 

T

0

 = 

T

1

 = 

T

2

From the period duration 

T

0

 measured in step 6.3 

for  a  calibrated  reversible  pendulum  and  a  dis-
tance 

l

 = 0.8 m  between  the  two  fulcra,  corre-

sponding to the shortened pendulum length, it is 
possible  to  determine  the  acceleration  due  to 
earth’s gravity: 

2

2

0

4

l

g

T

   

Note: To adjust the pendulum so that  both peri-
ods of oscillation are exactly the same, it may be 
necessary  to  turn  the  pendulum  bob  by  180° 
about the vertical and mount it on the pendulum 
rod that way round. 

7. Storage, cleaning, and disposal 

 

Keep the apparatus in a clean, dry, dust-free 
place. 

 

Do not use aggressive cleaning agents or sol-
vents to clean the apparatus. 

 

In  order  to  clean  the  equipment,  use  a  soft, 
damp cloth. 

 

The packaging should be disposed of at local 
recycling centres 

 

If the equipment itself is 
to  be  disposed  of,  it 
must  not  be  included 
with  normal  household 
waste.  Local  regula-
tions are to be obeyed. 

 

 
 

 

Fig. 6: Period of oscillation 

T

 as a function of measuring point number 

n

. Red circles: Both pendulum bobs below 

fulcrum. Black squares: Fixed pendulum bob above fulcrum, movable bob below

 

Summary of Contents for 1018466

Page 1: ... up the stand on solid level ground and fasten the assembly screws tightly Fasten the securing screw for the movable pendulum bob tightly so that the bob cannot move unchecked Use the adjustment screws in the base plate of the stand to align the bearing plate in such a way that the bearing axes of the pendulum bob can be even Always handle the pendulum with both hands when setting it up or adjusti...

Page 2: ...endulum bob b Pendulum rod c Bearing fulcrum d Latch in positions e Movable pendulum bob f Securing screw g Assembly screws Stand h Stand board i Positioning screws j Stand rod k Stand board l Bearing plate m Spirit level n Floor cover ...

Page 3: ...ration between bearing fulcra 80 cm Fixed pendulum bob 1 4 kg approx Movable pendulum bob 1 0 kg approx Pendulum deflection max 10 cm 5 Set up 5 1 Choice of set up location If the floor on which the equipment is set is springy then the oscillation energy is transferred to the whole stand and measurement errors can then arise Only set up and use the reversible pendulum on solid level ground On smoo...

Page 4: ...blue are below their corresponding fulcra Attach the movable bob in the wedge shaped groove nearest to the fixed bob i e the low est one This position will subsequently be re ferred to as measuring point 1 even when the pendulum rod is suspended above the other fulcrum Measure the period of oscillation and make a note of it Keep moving the movable bob from one groove to the next and secure it meas...

Page 5: ... due to earth s gravity 2 2 0 4 l g T Note To adjust the pendulum so that both peri ods of oscillation are exactly the same it may be necessary to turn the pendulum bob by 180 about the vertical and mount it on the pendulum rod that way round 7 Storage cleaning and disposal Keep the apparatus in a clean dry dust free place Do not use aggressive cleaning agents or sol vents to clean the apparatus I...

Reviews:

Related manuals for 1018466