background image

 

3. Technical data 

Franck-Hertz tube 

Filament: 

4 to 9 V AC/DC 

Grid voltage: 

0 to 80 V 

Bias voltage: 

1.5 V approx. 

Operating temperature:  160°C - 200° C approx. 
Tube dimensions: 

160 mm x 30 mm diam. 

Weight: 

380 g approx. 

Heating chamber 

Mains voltage: 

 See back of case 

 

Dimensions of front  

opening: 

230 x 160 mm² approx. 

Heating power: 

800 W @230 V 

 

400 W @115 V 

Maximum temperature:  300°C @230 V 

250°C @115 V 

Temperature constancy:  ±1°C approx. 

Dimensions: 

335x180x165  mm³  ap-
prox. 

Weight: 

5.6 kg approx. 

 

4. Operation 

The following equipment is also required to com-
plete the experiment: 

1 Power supply unit for F/H experiment @230 V 

   

1012819 

or 
1 Power supply unit for F/H experiment @115 V 

   

1012818 

1 Digital oscilloscope, 2x 30 MHz 

1020910 

1 HF Patch cord, 1 m 

1002746 

2 HF Patch cords, BNC/4 mm plug 

1002748 

Safety leads for experiments 

1002843 

 

 

Note: Before switching on, remove the PE 
packaging  part,  which  is  located  behind 
the  tube,  from  the  inside  of  the  heating 
oven. 

 

Place front plate of the open side of the heat-
ing chamber and fix it in place with 6 knurled 
screws. 

 

Turn off the heating chamber and the control 
unit to begin with and turn all the knobs on the 
control fully to the left. 

 

Do  not  apply  an  accelerating  voltage  to 
the tube when it is still cold (the mercury 
inside may cause a short circuit). 

 

Connect terminals "A", "F" and "K" (refer to fig. 
2).  

 

Connect  terminal  "E"  of  the  Franck-Hertz 
tube to the correct input on the control unit by 
means of an BNC cable.  

 

Connect  the  “FH  Signal 

U

Y

-

out”  terminal  of 

the  control  unit  to  the  Y  input  of  the  oscillo-
scope and terminal “

U

X

” to the X input. 

 

Turn  on  the  control  unit  and  the  equipment 
should enter ramp mode. 

 

Slowly set a filament voltage of 6 V. The indi-
rectly heated cathode requires about 90 sec-
onds to warm up, once the voltage is applied. 

 

Turn on the heating chamber. Set a tempera-
ture of about 180° C and wait for the tube to 
warm up (about 5 to 10 minutes). 

 

Set  the  minimum  acceleration  voltage  to 
zero, slowly increase the maximum accelera-
tion voltage to 80 V. 

 

Do  not,  however,  increase  the  accelerating 
voltage so much that self-discharge no longer 
occurs inside the tube. Any ionisation due to 
collisions will disrupt the curve. 

 

Set up the oscilloscope initially  with settings 
of x = 1 V/div and y = 1 V/div. 

 

Observe the emergence of the maxima in the 
Franck-Hertz  trace  on  the  oscilloscope 
screen. 

 

Set up  all  the  parameters,  accelerating  volt-
age, cathode filament, bias voltage and am-
plitude so that a trace with nicely delineated 
maxima and minima is obtained. 

The  procedure  as  described  so  far  is  a  gen-
eral  setting  procedure.  Unavoidable  differ-
ences resulting from the manufacture of indi-
vidual  Frank-Hertz  tubes  mean  that  the  opti-
mum parameters may differ from tube to tube. 
The test report included with the tube should 
give some idea of where good results may be 
obtained for the tube in question. 

The collector current displays regularly recurring, 
equidistant  maxima  and  minima  that  are  inde-
pendent  of  the  accelerating  voltage.  The  interval 
between these peaks is 4.9 V. A contact potential 
of  2 V  exists  between  the  anode  and  cathode  of 
the tube, which is why the first maximum only ap-
pears in the region of 7 V. The first maxima will be 
more obvious when the temperature of the heat-
ing chamber is lower. 

 

Evaluation of the Franck-Hertz curve: 

To fully evaluate the Franck-Hertz curve, a digital 
voltmeter  is  needed.  This  does  not  require  that 
the  current  of  the  electron  beam  be  determined 
precisely.  The  oscilloscope  screen  should  show 
the trace  of a Franck-Hertz curve featuring very 
clear maxima and minima. 

Summary of Contents for 1023095

Page 1: ...d personnel Before putting the equipment into operation con firm it is compatible with the local mains voltage Before setting up or starting any experiments check the apparatus for any damage In the event of any malfunction defect or visi ble damage switch off the equipment imme diately and secure it against any inadvertent use The instrument may only be connected to the mains via a socket that ha...

Page 2: ...ered alumina ring A fixed attenuating resistor 10 is inserted between the sockets for the accelerating voltage and for the anode of the tube This protects the tube in the event of excessive voltage occur ring due to arcing The voltage drop across this re sistor can be neglected when measurements are being made Heating chamber The heating chamber serves to establish the va pour pressure inside the ...

Page 3: ...indi rectly heated cathode requires about 90 sec onds to warm up once the voltage is applied Turn on the heating chamber Set a tempera ture of about 180 C and wait for the tube to warm up about 5 to 10 minutes Set the minimum acceleration voltage to zero slowly increase the maximum accelera tion voltage to 80 V Do not however increase the accelerating voltage so much that self discharge no longer ...

Page 4: ... accelerating voltage must be lowered immediately If it should be necessary to map further maxima the temper ature of the furnace must be increased before hand Optimization of the countervoltage Figures 5 b and 5 c show a too steep or too flat curve with few maxima The countervoltage de termines the slope of the curve The greater the countervoltage the flatter the rise In combination with the acce...

Page 5: ... regulations for the disposal of elec trical equipment will ap ply If Frank Hertz tubes are to be disposed of local regulations applying to the disposal of mercury must be followed Fig 1 Schematic of set up for measuring the Franck Hertz curve for mercury C cathode G grid A collector elec trode Fig 2 Experiment set up for Franck Hertz tube with Hg filling F K G A E X Y E F K A ...

Page 6: ...Str 20 20459 Hamburg Germany www 3bscientific com Subject to technical amendments Copyright 2021 3B Scientific GmbH Fig 3 Experiment set up for Franck Hertz tube with 2 digital voltmeters Fig 4 Franck Hertz curve F K G A E E F K A X Y ...

Reviews: