background image

 

 

Connect a digital voltmeter between the sig-
nal output (

U

X

) and the ground socket (refer 

to fig. 3). 

 

Press the “Man/Ramp” button and the display 
will show “Man” to indicate manual mode. 

 

Turn  the  accelerating  voltage  knob  all  the 
way to the left (

U

A

 = 0 V). 

The display will show the accelerating voltages in 
steps  of  0.5  V.  In  order  to  obtain  better  test  re-
sults,  a  digital  voltmeter  can  be  connected  be-
tween  sockets  "A"  and  "K"  in  order  to  obtain  a 
more accurate measure of the accelerating volt-
age. 

Note:  The  accelerating  voltage  is  reduced  by  a 
factor of 10 at the signal output (

U

X

). The digital 

voltmeter, however, measures the full accelerat-
ing voltage 

between sockets “A” and “K”. 

By gradually increasing the accelerating voltage 
at  a  constant  rate,  the  precise  position  of  the 
maxima and minima can be determined with the 
aid of the digital voltmeter. 

 

 
 
 

Curve optimization 

 

 

 

Fig. 5: Illustration of various signals of the Franck-
Hertz tube. 

 

The  parameters  of  the  tube  can  change  during 
the use of the tube. This usually happens either 
due  to  ageing  of  the  tube  or  due  to  the  experi-
mental  specification.  Figure 

5.f) 

represents  the 

ideal example of the curve. However, other types 
of curves can also occur. 

 

Breakdown of the tube: 

 

Figure 

5.a)

 shows the case of a breakdown of the 

tube.  The  anode  current  increases  rapidly  to  a 
maximum  value.  In  this  case,  the  accelerating 
voltage must be lowered immediately. If it should 
be necessary to map further maxima, the temper-
ature  of  the  furnace  must  be  increased  before-
hand. 

 

Optimization of the countervoltage: 

 

Figures 

5.b)

 and 

5.c)

 show a too steep or too flat 

curve  with  few  maxima. The  countervoltage  de-
termines the slope of the curve. The greater the 
countervoltage, the flatter the rise. In combination 
with  the  acceleration  voltage,  the  quality  of  the 
maxima can be slightly improved. 

 

Optimization of the filament voltage: 

 

The  filament  voltage  determines  the  number  of 
emitted  electrons  and  thus  the  anode  current. 
The  higher  the  filament  voltage,  the  more  elec-
trons  are  emitted.  Figure  5.d)  shows  the  case 
where  the anode current  is too high. The signal 
flattens out to a horizontal line above a threshold 
value.  The  maximum  number  of  available  elec-
trons  is  reached  (under  the  given  filament  volt-
age) and the anode current remains constant, de-
spite the increasing accelerating voltage. Figure 
5.e) shows the case of a too low filament voltage. 
The  signal  has  a  low  rise  and  the  maxima  are 
weak, although there is a sufficiently high number 
of maxima.  In these cases, lowering or increas-
ing the filament  voltage  is  often sufficient to ob-
tain  an  evaluable  curve.  Note:  An  excessively 
high filament voltage has a negative effect on the 
life of the tube. It is not recommended to operate 
the tube with too high filament voltages. 

 

 

5. Care and maintenance 

 

Before cleaning the equipment, disconnect it 
from its power supply. 

 

Use a soft, damp cloth to clean it. 

 

Summary of Contents for 1023095

Page 1: ...d personnel Before putting the equipment into operation con firm it is compatible with the local mains voltage Before setting up or starting any experiments check the apparatus for any damage In the event of any malfunction defect or visi ble damage switch off the equipment imme diately and secure it against any inadvertent use The instrument may only be connected to the mains via a socket that ha...

Page 2: ...ered alumina ring A fixed attenuating resistor 10 is inserted between the sockets for the accelerating voltage and for the anode of the tube This protects the tube in the event of excessive voltage occur ring due to arcing The voltage drop across this re sistor can be neglected when measurements are being made Heating chamber The heating chamber serves to establish the va pour pressure inside the ...

Page 3: ...indi rectly heated cathode requires about 90 sec onds to warm up once the voltage is applied Turn on the heating chamber Set a tempera ture of about 180 C and wait for the tube to warm up about 5 to 10 minutes Set the minimum acceleration voltage to zero slowly increase the maximum accelera tion voltage to 80 V Do not however increase the accelerating voltage so much that self discharge no longer ...

Page 4: ... accelerating voltage must be lowered immediately If it should be necessary to map further maxima the temper ature of the furnace must be increased before hand Optimization of the countervoltage Figures 5 b and 5 c show a too steep or too flat curve with few maxima The countervoltage de termines the slope of the curve The greater the countervoltage the flatter the rise In combination with the acce...

Page 5: ... regulations for the disposal of elec trical equipment will ap ply If Frank Hertz tubes are to be disposed of local regulations applying to the disposal of mercury must be followed Fig 1 Schematic of set up for measuring the Franck Hertz curve for mercury C cathode G grid A collector elec trode Fig 2 Experiment set up for Franck Hertz tube with Hg filling F K G A E X Y E F K A ...

Page 6: ...Str 20 20459 Hamburg Germany www 3bscientific com Subject to technical amendments Copyright 2021 3B Scientific GmbH Fig 3 Experiment set up for Franck Hertz tube with 2 digital voltmeters Fig 4 Franck Hertz curve F K G A E E F K A X Y ...

Reviews: