5

 

7. Current measurements up to 3 A 

 

The ESCOLA 100 may NOT be used 
to measure circuits connected directly 
to  low-voltage  mains  sources.  See 
page  2  of  “Measurement  categories” 
as specified in DIN EN 61010-1. 

 

The  nominal  voltage  of  the  current 
source may not exceed 600 V. 

 

The multimeter is to be connected 
in series with the load on the side 
where the voltage  with respect to 
ground is lowest! 

 

Turn off the circuit before the Es-
cola  100  meter  is  connected  into 
it. 

 
 

 

 
 

 

If measurements are made when the magni-
tude of the variable is unknown, always select 
a large measuring range before shifting down 
to lower ones. 

 

Connect the terminal at the lower potential to 
the common/ground socket. 

 

Connect  the  common/ground  lead  first  and 
only then the signal lead. 

 
 

7.1  DC currents up to 3 A: 

 

Set the four-way switch to 

 

Alternatively, if measurements are to be made 
with the zero point in the centre, then set the 
four-way switch to 

 

Select  the  required  current  measuring  range 
to a range measured in mA or A. 

 
 

7.2  AC currents up to 3 A: 

 

Set the four-way switch to 

 

Select  the  required  current  measuring  range 
to a range measured in mA or A. 

 

8. Current measurements up to 10 A 

 

The ESCOLA 100 may NOT be used 
to measure circuits connected directly 
to  low-voltage  mains  sources.  See 
page  2  of  “Measurement  categories” 
as specified in DIN EN 61010-1. 

 

The  nominal  voltage  of  the  current 
source may not exceed 600 V. 

 

The multimeter is to be connected 
in series with the load on the side 
where the voltage  with respect to 
ground is lowest! 

 

Turn off the circuit before the Es-
cola  100  meter  is  connected  into 
it. 

 
 

 

 
 

 

Set up a measuring range of 10 A. 

 

Connect the terminal at the lower potential to 
the common/ground socket. 

 

Connect  the  common/ground  lead  first  and 
only then the signal lead. 

 
 

8.1  DC currents up to 10 A: 

 

Set the four-way switch to 

 

Alternatively, if measurements are to be made 
with the zero point in the centre, then set the 
four-way switch to 

 
 

8.2  AC currents up to 10 A: 

 

Set the four-way switch to 

 
 
 
 
 
 
 
 
 

~

~

10A

V

A,

CAT III 600V

~

~

10A

V

A,

CAT III 600V

Summary of Contents for Escola 100

Page 1: ...C quantities and for continuity tests featuring an acoustic signal AC quantities with frequencies of up to 40 kHz can be measured All the possible measuring ranges can be selected by means of a single rotary switch The device is protected by a fuse and authorised for measurements on wiring installations in CAT III e g in distribution boxes consumer units The current measuring ranges are resistant ...

Page 2: ...igh frequency circuits where dangerous voltages may arise due to superimposition of components The authorised measuring range is not to be exceeded If measurements are made when the magnitude of the variable is unknown al ways select a large measuring range before shifting down to lower ones Make very sure that the voltage value be tween the measured contact and earth or be tween the ground socket...

Page 3: ...mined by forming various quotients involving simultaneous meas urements of current and voltage R U I below 1 mΩ above 10 MΩ S I U below 1 µS above 30 S Z U I below 1 mΩ above 10 MΩ 40 Hz 40 kHz Y I U below 1 µS above 30 S 40 Hz 40 kHz Overload protection Voltage ranges 600 V long term in all voltage ranges Current ranges 10 A of long term load ing in 3 A and 10 A ranges Electrical safety Safety sp...

Page 4: ...easuring leads Set the rotary switch to If the battery is sufficiently charged the needle will point to the following range indication If this is not the case the battery will need to be re placed immediately 6 3 Zero point calibration Set up the analogue multimeter either hori zontally or vertically Set the four way switch to Turn the rotary switch to 600 V Connect the common ground socket and th...

Page 5: ...itch to Select the required current measuring range to a range measured in mA or A 7 2 AC currents up to 3 A Set the four way switch to Select the required current measuring range to a range measured in mA or A 8 Current measurements up to 10 A The ESCOLA 100 may NOT be used to measure circuits connected directly to low voltage mains sources See page 2 of Measurement categories as specified in DIN...

Page 6: ... component to be tested Set the four way switch to and turn the rotary switch to If there is continuity of cur rent R 1 kΩ an audible signal will sound 11 Measuring current and voltage together The ESCOLA 100 may NOT be used to measure circuits connected directly to low voltage mains sources See page 2 of Measurement categories as specified in DIN EN 61010 1 The nominal voltage of the current sour...

Page 7: ...the casing again 12 3 Changing fuses Before opening the casing Turn off the Escola 100 Disconnect all measuring leads Unscrew the back of the casing Check the fuses Blown fuses should be replaced with ones of the same rating Close the casing again 13 Cleaning For cleaning use a soft cloth slightly mois te ned with alcohol or a brush Electrostatic charging of the display window can affect the measu...

Page 8: ......

Reviews: