Luna G3 BPAP 25A (G3 B25A) User Manual MID-97/v1.0
32 / 48
17. Alert
Alert Message
Description
Power Failure!!!
An audible alert will sound in 6s if the device is accidentally
disconnected from power when it is delivering air.
Note:
(1) The alert will not sound if power failure occurs when the
device is in standby state.
(2) No alert message displayed during a power failure.
Device fault!!!
An audible alert will sound if no airflow comes out of the
machine; the screen will display “
Device fault!!!
”.
Leak!!
When the airflow is on, an audible alert will sound if the air
leak rate exceeds 150 L/min; the screen will display
“
Leak!!
”.
Low Input
Voltage!!
If the voltage supplied by power adaptor is lower than
22V, an audible alert will sound and the screen will display
“
Low Input Voltage!!
”.
Humidifier
Failure!!
When humidifier is connected, an audible alert will sound if
the humidifier fails to work in 10 minutes; the screen will
display “
Humidifier Failure!!
”.
Please Change
Filter!
When the Filter Alert feature is enabled, an audible alert
will sound if the preset replacement time is reached but
the air filter hasn’t been replaced; the screen will display
“
Please Change Filter!
”. The default setting is “
Off
”.
Please Change
Tube!
When the tubing Alert feature is enabled, an audible alert
will sound if the preset replacement time is reached but
the tubing hasn’t been replaced; the screen will display
“
Please Change Tube!
”.
Please Change
Mask!
When the Mask Alert feature is enabled, an audible alert
will sound if the preset replacement time is reached but
the mask hasn’t been replaced; the screen will display
“
Please Change Mask!
”.
SD Card Full!
The screen will display “
SD Card Full!
” if the SD card has
reached its maximum capacity.
Reinsert SD
card!
The screen will display
“
Reinsert SD card!
” if
the SD card
fails to work.