Luna G3 BPAP 25A (G3 B25A) User Manual MID-97/v1.0
30 / 48
16.1.7 Exiting the Patient Menu
The users can press
the Home
to return to the Main Interface shown in Fig. 16-1.
16.2 Options of the Patient Menu and Corresponding
Descriptions
Option
Range
Description
Humidifier
Off,Auto,
1
~
5
There are five humidity levels available. As the
numbers
increase,
the
humidity
rises
accordingly. “Off” means the humidifier is
turned off.
Preheat
On / Off
Set humidifier to preheat by adjusting this
option. This feature automatically turns off after
30 minutes.
Reslex
Off, 1
~
3
This feature enables the device to automatically
reduce the treatment pressure when the patient
exhales, so as to make the user more comfortable.
The higher the number is, the more pressure the
device reduces. “Off” means this feature is
disabled.
Heated
Tubing
Off, 1
~
5
There are five heat levels available. As the
number increases, the heat rises accordingly.
“Off”
means the heat is turned off. Note:
Heated Tubing
is displayed in the patient menu
only when it is connected.
Ramp Time
Auto, 0 -
Max Ramp
In order to increase comfort and help the
patient fall asleep easily, the pressure can
increase graduallywhen the Ramp feature is
enabled. The ramp time during which the initial
pressure rises to the prescribed treatment
pressure can be adjusted. As you turn
the Knob
to the nearest point, the numbering
increases or decreases by five minutes. The
screen displays a real-time countdown of the
remaining ramp time in seconds.
Delay
On / Off
When the humidifier is on, this feature allows the
airflow to continue for about 15 minutes at a
low pressure (about 2 cmH
2
O) after you press
the Start / Stop
to discontinue
treatment. This will blow off the vapor left in the
water chamber to avoid any damage to the
device. When this feature is set to “Off,” which
means it is disabled, the airfolw stops delivering
air instantly after you press
the Start / Stop
.
Date
2000-01-01
—
2099-12-31
Set date by adjusting this option.