6. Specifications
Device Size
Dimensions (L x W x H): 265 mm × 145 mm × 114 mm
Weight: 1.7 kg
Water capacity:
To maximum fill line 360 mL
Product Use, Transport and Storage
Operation
Transport and Storage
Temperature: 5°C to 35°C
(41˚F to 95˚F)
-25°C to 70°C (-
13˚F to 158˚F)
Humidity: 15% to 93% Non-condensing 15% to 93% Non-condensing
Atmospheric Pressure: 760
~
1060 hPa 760
~
1060 hPa
Heated Humidifier
Humidifier Settings: off, Aut
o, 1 to 5 (95˚F to 154.4˚F
/ 35°C to 68°C)
Humidifier Output: No less than 15 mg H
2
O/L
Environmental Conditions: Maximum airflow, 35°C, 15% relative humidity
Maximum Operating Pressure: 40 cmH
2
O
Pressure Drop with Humidifier: < 0.4 cmH2O at 60 LPM flow
Maximum Delivered Gas Temperat
ure: ≤ 43°C
Mode of Operation
Continuous
Work Mode
CPAP, AutoCPAP, S, AutoS
SD Card
The SD card can record patient data and fault information
AC Power Consumption
100V - 240 V
~
, 50Hz / 60 Hz, 2 A
Device offer to Heated Tubing Communications Port
24 V
18 W
Type of Protection Against Electric Shock
Class II Equipment
Degree of Protection Against Electric Shock
Type BF Applied Part
Degree of Protection Against Ingress of Water
IP22
Pressure Range
4 to 25 cmH
2
O
(in 0.5 cmH
2
O increments)
Under single fault conditions,
≤
30 cmH
2
O for CPAP and AutoCPAP mode,
≤
40 cm
for the rest of the modes.