Pannoramic 1000 2.0 – User’s Guide
4 Software and Work-flow
◦
ICC Profile
– Select an applicable preset color profile to be used for Brightfield scanning
from the drop-down menu. Settings according to the profile has effect on the color set,
color temperature, and gamma value of the scanned image.
Preset profiles
are the
following:
▪
Calibrated Linear
: Color scheme is defined by calibration (Gamma value is not
included)
▪
sRGB standard
: Color scheme is defined by calibration (standard Gamma value
-2.2- is applied)
▪
sRGB standard Microscope Like
: Color scheme is defined by calibration
(standard Gamma value / White Point: D50)
▪
Custom
: parameters are freely customizable as follows:
•
Color Scheme
◦
Calibrated
: scheme is defined by the color calibration slide
◦
Raw
: the color information of the recorded raw image is applied
◦
Custom
:
White Point
,
Gamma
, and
Color Matrix
can be set for your own
requirements
•
White Point
◦
D65
: color temperature according to a standard illuminant D65 white
point value is applied (6500K / daylight)
◦
D50
: color temperature according to a standard illuminant D50 white
point value is applied (5000K / horizontal daylight)
•
Gamma
◦
Standard
: a value of 2.2 is used
◦
Medium
: a value of 1.8 is used
◦
Off
: a value of 1.0 is used
•
Custom Color Matrix
NOTE:
If you do not have extensive experience with color correction settings,
then modifying the values in the matrix is not recommended.
May 23, 2019 - Rev. 1