Pannoramic SCAN 150 1.15 User’s Guide
4 Software Description
•
If modification is done to the folder tree in the CaseCenter server, click
Update Tree
to refresh the tree view.
•
Click
Done
when finished or
Cancel
to quit the CaseCenter window. If you save the
form, slides will be scanned to server according to the data recorded in the form
after you start scanning.
•
After restarting the scanner software a dialog window appears when scanning to
server is performed previously. Click
Yes
to use the last used server, or
No
to define
a different one.
4.8 Using Profiles
Note:
Saved profile settings cannot be transferred to and used on another
control computer.
If you want to scan a group of slides, it is advised to save your settings previously in a profile and
continue scanning in Automatic mode. You can save the following settings in a profile:
Scan Mode
•
Brightfield
•
Fluorescent
Scan Options
•
Focus frequency (present in Fluorescent mode)
•
Interpolated Focus distance (present in Brightfield mode)
•
Compression procedure
•
Use of a fixed compensation image (ON/OFF)
•
Use of stitching (ON/OFF)
2012. March 13.
3DHISTECH Ltd.
107
(