Pannoramic SCAN 150 1.15 User’s Guide
5 Troubleshooting and Maintenance
Virtual slides show poor stitching in Pannoramic Viewer
•
Scan camera not properly adjusted (for example, due to loose fixture)
•
Contact 3DHISTECH Service for support.
Slide is often positioned incorrectly
•
Guiding fault in transport system
•
Check if faulty positioning disappears after the selection of a different type of
magazine (for example, universal magazine).
•
Contact 3DHISTECH Service for support.
Digitization process is extremely slow
•
Extended focusing
option is activated
•
Deactivate the
Extended focusing
option if not required.
•
Z-stack
option is activated
•
Deactivate the
Z-stack
option if not required.
Image is out of focus (in Manual mode)
•
Focus range limit is activated
•
Deactivate the
Limiting focus range
option.
•
Focus range setting is incorrect
•
Change focus range setting.
All individual images show a shading effect
•
Algorithm failed to find empty image fields
•
Capture a reference image with a blank slide (see section
4.3.2 Scan Options
).
March 13, 2012 - Rev. 1
3DHISTECH Ltd.
113(132)