Pannoramic SCAN 150 1.15 User’s Guide
3 Operating Slides and Magazines
3 Operating Slides and Magazines
3.1 Preparing Slides and Magazines
This section demonstrates how to prepare slides and magazines.
You can use all slides and cover slips that meet the following specifications:
Slides
Cover slips
Length
75.0 to 76.2 mm
max. 50 mm
Width
25.0 to 26.0 mm
max. 24 mm (recommended: 22 mm)
Thickness
0.95 to 1.05 mm
No. 1 and No. 1.5 (0.13 to 0.16 mm and 0.16 to 0.19
mm)
Corners
45° beveled corners
Edges
Grounded or cut
Note:
FISH fluorescent signals become more easily detectable for the
scanner, therefore resulting in a brighter and better image, by treating
the slide with a formula that prevents rapid photobleaching of
fluorescent reagents (e.g. VECTASHIELD® Fluorescent Mounting
Medium by
Vector Laboratories
and ProLong® Gold and
SlowFade
®
Gold antifade reagents by
Invitrogen
) during the FISH protocol.
For further information, see the homepages of the manufacturers:
and
Warning!
Do not use a slide when it is broken or its edges are not properly cut or
grounded. Using such slides can lead to injury.