background image

3D Systems, Inc.

6

p/n 76-D022 Rev. A

ProX

®

 SLS 6100 3D Printer

Supplied by

: 3D Systems

Installed by

: 3D Systems

The printer requires a 3-phase, 208 VAC, 50/60 Hz, 10 kVA power source .
Ceiling drop lines for nitrogen supply and exhaust, electrical, and clean dry air (CDA) are recommended. The floor below the 

process station must be flat and vibration-free. The room must be air-conditioned, with no vents directly above the machine. 

Clearance around the process station is required so hinged access panels can be fully opened . See the 

ProX

®

 SLS 6100 3D 

Printer Facility Requirements Poster in section 14 of this Facility Guide

 

for clearance requirements .

F

B

A

D

E

C

Stacklight: 

Indicates the state of the 

system .

Print Chamber Area: 

There are two doors 

in front of the Print Chamber—the outer 

locking door, and the inner print chamber 

door . The parts are printed inside the Print 

Chamber .

E-Stop: 

The Emergency Stop button is a 

safety mechanism used to shut off the 

machine in an emergency situation in which 

it cannot be shut down in the usual 

manner .

User Interface (UI):

 The user controls the 

system using a monitor located on the front 

panel .

LED Interface Lights and Controls:

 There 

are three LED’s which give the user visual 

feedback regarding the system . The two 

buttons allow the user to control the 

chamber lights and enable the system . 

There is (1)  standard USB ports beneath 

the LED’s . .

User and Service Access Panels: 

There are 

two types of panels on the sides and back 

of the machine—user access and service 

access . The user access panels are hinged 

doors . The service access panels are lift-off 

panels . A tool is required to open each 

service panel .

A

B

C

D

E

F

Summary of Contents for ProX SLS 6100

Page 1: ...3D Systems Inc ProX SLS 6100 3D Printer Facility Guide Original Instructions...

Page 2: ...ENTS CHECKLIST 13 About the Checklist 13 How to Submit Your Completed Facilities Checklist 13 Facility Requirements Checklist ProX SLS 6100 3D Printer and Auxiliary Equipment 14 7 ROOM REQUIREMENTS 15...

Page 3: ...Rev A 11 NETWORK REQUIREMENTS 20 12 SHIPPING WEIGHTS AND DIMENSIONS 21 System Component Weights and Dimensions 21 13 PROX SLS 6100 SHIP KIT 22 14 SINGLE MQC SHIP KIT 23 15 ProX 6100 SLS 3D PRINTER FA...

Page 4: ...nter is crated stored transported and un crated are covered in this Facility Guide available to the user This guide contains a described sequence for crating re crating and transport Details on liftin...

Page 5: ...l grab the handle and pull it towards you to pop out the first clip NOTE To keep clips from popping off too quickly you may want to hold your hand over the top of the clip 2 Using at least a 5 foot si...

Page 6: ...rner of the printer 7 Save the 4 screws and brackets for future crating use 8 Using a forklift capable of lifting 3000 pounds spread the forks from the forklift so they span the width of the open ing...

Page 7: ...owly lower the printer to the floor 13 Roll the printer to the desired destination NOTE If recrating the printer please use the instructions in reverse order 14 Once the printer has been rolled into t...

Page 8: ...rder These are indicated in each section The facilities for these non 3D Systems supplied components and in some cases the components themselves must be in place before the ProX SLS 6100 SLS Single MQ...

Page 9: ...are two doors in front of the Print Chamber the outer locking door and the inner print chamber door The parts are printed inside the Print Chamber E Stop The Emergency Stop button is a safety mechani...

Page 10: ...powder ratios The tag must be scanned to unlock the Fresh Material input doors and add the material to the system Breakout Area The print cake is brought here using the Print Transfer Cart Assembly R...

Page 11: ...tor for Oxygen Sensor Nitrogen Inlet Input for nitrogen supply Quick Disconnect Air Fitting Input for compressed air coming into the MQC System Pressure Regulator Nitrogen Pressure Regulator Air Press...

Page 12: ...ylinder Part Transfer Cart Supplied by 3D Systems Installed by 3D Systems The print cake Extraction Cylinder Part Transfer Tray and Part Transfer Cart are used to remove the print cake from the ProX S...

Page 13: ...ier 3D Systems or customer The Dust Extractor 230VAC 60Hz 3PH and 400VAC 50Hz 3PH draws airborne powder out of the MQC System during part breakout through a flexible vacuum duct and collects airborne...

Page 14: ...ded in the ProX SLS 6100 User Guide for information on safely working with nitrogen and oxygen BEAD BLASTER Supplied by 3D Systems Installed by 3D Systems If you plan to use DuraForm SLS material 3D S...

Page 15: ...o you can properly size your tanks SLS MATERIALS The material listed below is available for purchase from 3D Systems The material comes with its own 3D Systems Material Guide to help you use it succes...

Page 16: ...Poster for example when you lay out the room or wire electrical power The instructions tell you when you need to refer to this drawing After your facility meets all the requirements for a section chec...

Page 17: ...ed to complete the installation Contact name Phone email fax Phone Email Fax Facility address Date submitted Room Requirements completed Atmosphere Requirements completed Electrical Requirements compl...

Page 18: ...IGHT Room 481 cm 16 ft 427 cm 14 ft 305 cm 10 ft Access door1 290 cm 9 5 ft 260 cm 8 5 ft 1 The SLS Single MQC System is 244 cm 8 ft tall and 275 cm 9 ft wide but extra clearance is needed if it is on...

Page 19: ...IABLES REQUIREMENTS Room temperature controls Heating and air conditioning installed A C not blowing on top of process station Temperature Operating range 16 32 C 60 to 90 F Stability 2 C 5 F Non cond...

Page 20: ...cility meets has met its required local electrical code NOTE The ProX SLS 6100 is designed to be connected to primary AC power directly from the facility s power circuit to the machine s input power l...

Page 21: ...g current 4A On Off breaker rating 5A Power Cable IEC320 C19 Plug 16AWG 1 29mm 2 Conductors P E Connection Maximum Room Ambient Temperature 27 C GROUNDING REQUIREMENTS All connections between powered...

Page 22: ...y and O2 waste Both exit at the bottom of the generator The N2 supply stream is greater than 99 nitrogen The O2 waste stream is less than 40 oxygen If the SLS process room meets the air exchange requi...

Page 23: ...s messages There is one ethernet connector on the ProX SLS 6100 system that is connected to the SLS Single MQC and another that is connected to the facility s LAN The ethernet connectors are RJ 45 usi...

Page 24: ...ND DIMENSIONS Component Weight2 Width Depth Height Printer Station1 1360 kg 174 cm 123 cm 230 cm 3000 lb 69 in 48 in 90 in SLS Single MQC System 800 kg 238 cm 99 cm 236 cm 1764 lb 93 9 in 39 in 93 in...

Page 25: ...alibration 1 132349 00 Cable O2 ESR Alarm 50 Foot DDH 1 132350 00 Cable O2 ESR Alarm Jumper DDH 1 132572 00 Lens Tissue 1 132896 00 CAT5E SHIELDED CABLE 50 ft GRAY 1 23099 107 00 DRIVER TORXPLUS STR I...

Page 26: ...ression 75 x 75 MPT Brass 1 106299 00 Sinterstation Pro 3 4 OD Tube Compression Sleeve 1 132350 00 Cable O2 ESR Alarm Jumper DDH 1 132725 00 KIT PROX 500 DUST COLLECTOR 1 132750 00 PROX MQC WEIGHT ERR...

Page 27: ...circuit breaker wiring REQUIREMENTS 208 VAC 10kVA 50 60 Hz 3PH 10 to 21 A 25 A 35 A 3 Delta Conductors P E connection phase 1 to L1 phase 2 to L2 phase 3 to L3 The 3 phase power cable and cable gland...