PDII123-UG-R

Pannoramic DESK II

1.23

User’s Guide

June 4, 2018

Rev. 2

For research and education uses only, not for use in diagnostic procedures.

Summary of Contents for Pannoramic DESK II

Page 1: ...PDII123 UG R Pannoramic DESK II 1 23 User s Guide June 4 2018 Rev 2 For research and education uses only not for use in diagnostic procedures ...

Page 2: ...Installation 18 2 1 Control computer ports 18 2 2 Connectors and cables 19 3 Preparing Slides 20 3 1 Affixing Barcode Stickers to Slides 21 3 2 Inserting and Removing Slides 22 4 Software Description 29 4 1 Starting Exiting Pannoramic DESK II Software Application 29 4 1 1 Starting the software 29 4 1 2 Exiting the software 29 4 2 Options 29 4 2 1 Barcode Settings 29 4 2 2 Image Compensation 34 4 2...

Page 3: ...67 5 1 2 Preventive maintenance 67 5 2 Troubleshooting 68 5 3 Maintenance 70 5 3 1 Removing the protective cover 70 5 3 2 Removing a slide 71 5 3 3 Transporting Pannoramic DESK II 73 6 Technical Data 74 Appendix Packaging instructions 77 Preparations 77 Main parts and their accessories 77 Packaging 78 Packing the DESK II scanner 80 Packing the accessories 80 Closing the package 80 Handling of the ...

Page 4: ...0344 US 11 662 976 HU0700409 US 12 042 050 US 77 636 144 and 007502032 The above mentioned granted patents granted utility model registered trademarks pending international and national patent applications and the other corresponding IP are the sole properties of the 3DHISTECH Ltd CAUTION For research and education uses only not for use in diagnostic procedures This product has not been approved o...

Page 5: ...3dhistech com Or contact the distributor Richard Allan Scientific Co Subsidiary of Thermo Fisher Scientific 4481 Campus Drive Kalamazoo Mi 49008 USA US Toll Free 1 800 522 7270 Tudor Road Manor Park Runcorn WA7 1TA United Kingdom 44 0 800 018 9396 44 0 1928 534 000 June 4 2018 Rev 2 3DHISTECH Ltd 5 82 For research and education uses only not for use in diagnostic procedures ...

Page 6: ...rements for Electrical Equipment for Measurement Control and Laboratory Use UL 61010 1 2004 R10 08 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use Part 1 General Requirements There is no further operator user action required in case of residual risks Further information may be obtained from the distributor Richard Allan Scientific Co Subsidiary of Thermo Fis...

Page 7: ...tion Example Cross references within this document or to other documents Warning It might cause irreversible damage or harm to the user of the product or the product itself if this instruction is not followed Caution It might cause the product to not work properly if this instruction is not followed and alerts to a potential danger situation Note Calls attention to some important features June 4 2...

Page 8: ...ized service personnel All claims for warranty and warranty services will be null and void in any such case Note The safety level of the equipment after servicing will be identical with level of the original equipment Warning Insert the power cord plugs only into a grounding outlet The protective action must not be rendered ineffective by extension cables without protective conductor The Pannorami...

Page 9: ...ing the lamp bulb Disconnect power plug and let the lamp cool down for about 15 minutes before replacing the lamp Caution Do not position the equipment so that it is difficult to operate and to disconnect the device if necessary Crush hazard Never open the front door unless prompted accordingly by software When you open the door at any other time make sure not to reach with your hands into the ins...

Page 10: ...ions necessarily vary to comply with differences in national and regional legislation Specific details can be found in the delivery documentation or from your dealer or representative Please note that your warranty may be invalidated if This instrument is modified in any way Accessories and reagents which have not been approved by Thermo Fisher Scientific are used The instrument is not operated or...

Page 11: ...of a thin glass plate on which specimens are mounted for microscopic study a dynamic interactive image that you can manage save magnify zoom name evaluate annotate mark and comment send to a colleague electronically for co operation or advice and so on June 4 2018 Rev 2 3DHISTECH Ltd 11 82 For research and education uses only not for use in diagnostic procedures ...

Page 12: ...c DESK II offers world class brightfield whole slide scanning An outstanding 0 12 µm pixel resolution is achieved with the 40 0 95 NA equivalent to 80 magnification Plan Apochromat objective Key features of the Pannoramic DESK II Scanner are Capable of scanning large slides 3MP camera for barcode reading and preview image Improved permanent auto focus Compact and robust device Good quality and sta...

Page 13: ...the labels used on the Pannoramic DESK II main unit and the external mains power supply unit Figure 1 Warning and information labels NOTE The symbol at the feeder slot calls attention to biohazardous materials in use June 4 2018 Rev 2 3DHISTECH Ltd 13 82 For research and education uses only not for use in diagnostic procedures ...

Page 14: ... Base Unit The following figure shows the main parts of the base unit Figure 2 Main parts of the base unit 1 Protective cover 2 Sliding door 3 Scan camera June 4 2018 Rev 2 3DHISTECH Ltd 14 82 For research and education uses only not for use in diagnostic procedures ...

Page 15: ... 2 1 Pannoramic DESK II system The following figure shows the main components of the Pannoramic DESK II system Figure 3 Pannoramic DESK II system overview 1 24 V external power supply unit 2 Pannoramic DESK II base unit 3 Monitor 4 Mouse 5 Control computer 6 Keyboard June 4 2018 Rev 2 3DHISTECH Ltd 15 82 For research and education uses only not for use in diagnostic procedures ...

Page 16: ...ightfield scanning Camera adapters 0 63 adapter 1 adapter Warning The installation of the cameras in Pannoramic DESK II must be carried out by trained professionals In any other case the warranty will be automatically voided The supplier or distributor is not liable for any damages caused by an installation carried out by an unauthorized person June 4 2018 Rev 2 3DHISTECH Ltd 16 82 For research an...

Page 17: ...ive if you need it to be set contact customer service 1 2 3 Control Software Pannoramic Scanner control software Optional licensed features Extended focus Z stack Barcode reader CaseViewer software fully functional Optional CaseCenter Pannoramic Viewer TMA module 3D Reconstruction module QuantCenter including applications such as HistoQuant NuclearQuant MembraneQuant and DensitoQuant modules Tumor...

Page 18: ... service assistants are allowed to install the hardware and software The product is ready for use after installation Caution Make sure that the device is placed so that there is enough space for ventilation and for safe and efficient operator maintenance 2 1 Control computer ports The following figure shows the location of the connectors in order to enable easier transportation of the product for ...

Page 19: ...nner and control computer Preview camera USB cable between scanner and control computer Point Grey Grasshopper 3 Monochrome USB 3 0 cable between Point Goint Grey Grasshopper 3 camera and control computer Video cable between monitor and control computer Power supply cable between 230V 110V and monitor Power supply cable between 230V 110V and control computer Power supply cable between 230V 110V an...

Page 20: ...us material or slides Warning Do not use a slide when it is broken or its edges are not properly cut or grounded Using such slides can lead to injury Warning Ensure that no part of the mounted coverslip protrudes from the edges of the slide Use slides with completely dried embedding medium only Otherwise the coverslip can be shifted that makes slide insertion less reliable Not completely dried emb...

Page 21: ...a 1 of the slide on the specimen side For large slides the label sticker must be applied on the left side of label area Caution Ensure that a marginal space of 1 2 mm remains on all sides between the sticker and the label area outer limits It is vital for barcode identification Do not let barcode sticker protrude from slide edges or stick on the coverslip surface 3 The thickness of the barcode sti...

Page 22: ...e slide type normal or large you are feeding into the device always make sure to ascertain that the adapter is applied or not How to place and remove the adapter For the placement and removal of the adapter a special tool is needed This applicator has a magnet at its head thus allowing an easy and fast placement or removal of the adapter Figure 6 Applicator for adapter June 4 2018 Rev 2 3DHISTECH ...

Page 23: ... door and locate the adapter marked with red on Figure 7 at the front edge of the specimen holder Figure 7 Location of the slide adapter 2 With the applicator gently remove the adapter from the specimen holder Figure 8 Lifting off the applicator 3 Put both the applicator and the adapter in a safe place June 4 2018 Rev 2 3DHISTECH Ltd 23 82 For research and education uses only not for use in diagno...

Page 24: ...device and the adapter is not placed on the specimen holder follow the below steps 1 Place the adapter to the head of the applicator 2 Open slide feeder door of the DESK II device 3 Place the adapter with the applicator onto the specimen holder Figure 10 Placing the adapter onto the specimen holder June 4 2018 Rev 2 3DHISTECH Ltd 24 82 For research and education uses only not for use in diagnostic...

Page 25: ...tor and make sure the adapter stays at its place Figure 11 Lifting off the applicator Note The device detects the presence of the adapter and therefore preview image area is automatically modified accordingly June 4 2018 Rev 2 3DHISTECH Ltd 25 82 For research and education uses only not for use in diagnostic procedures ...

Page 26: ... well trained person Figure 12 Inserting and removing a slide 1 Click in Pannoramic DESK II control software The specimen holder moves to the load unload position 2 Open the slide insert door 1 by pushing it to the left Caution When opening the slide insert door fingers may be pinched during operation 3 Open the fixing lever by pulling it to the left 4 Put the slide 3 you want to scan in the speci...

Page 27: ...well trained person 5 Close the fixing lever by pushing it to the right so that it touches and therefore fixes the slide 6 Close the slide insert door 1 by pushing it to the right Caution When closing the slide insert door fingers may be pinched during operation Warning Never open the front door unless prompted accordingly by software When you open the door at any other time make sure not to reach...

Page 28: ...e left 3 Open the fixing lever by pulling it to the left 4 Remove the slide 3 5 Insert another slide or close the fixing lever and the door 1 Note The slide insert door is equipped with an interlock switch It automatically stops a running digitization process when the slide insert door is opened June 4 2018 Rev 2 3DHISTECH Ltd 28 82 For research and education uses only not for use in diagnostic pr...

Page 29: ...rs the preview camera and the scan camera are initialized After successful initialization the program displays a Mode selection window Figure 13 Mode selection window 4 1 2 Exiting the software If you close the main window or one of the options window the Mode selection window turns active again To exit the software click X in the title bar of the Mode selection window Note that when exiting the s...

Page 30: ...Pannoramic DESK II 1 23 User s Guide 4 Software Description June 4 2018 Rev 2 3DHISTECH Ltd 30 82 ...

Page 31: ...ngs Barcode settings can be modified in the Barcode Settings window By predefining the barcode type foreground color black or white the color of the barcode and other settings you enable the Read Barcode option when using the control software Note June 4 2018 Rev 2 3DHISTECH Ltd 31 82 ...

Page 32: ...the slide is not tilted and that it rests perfectly horizontally on the stage To display an image of the barcode area click You can redefine the barcode reading area by dragging the edges of the green square To check whether the barcode reading settings are correct display the barcode digits in the Barcode field by clicking Note If you did not set the barcode type correctly or there is no barcode ...

Page 33: ...9 Code 128 Code 16K Interleaved 2 of 5 Standard 2 of 5 IATA 2 of 5 Matrix 2 of 5 Telepen Code 93 Code 11 MSI Codabar Patch Code Pharmacode EAN 13 EAN 8 UPC A UPC E 2D Barcode Types You can select the following 2D barcode types Aztec Data Matrix PDF417 Compact Micro PDF417 QR Micro QR MaxiCode June 4 2018 Rev 2 3DHISTECH Ltd 33 82 ...

Page 34: ...ick Use barcode compensation image to use the previously created compensation image for scanning the label area The compensation image improves the effectiveness of barcode reading 1 Prepare a slide with a label area covered fully by a white sticker 2 Take a new compensation image of the blank barcode sticker by clicking 4 2 2 Image Compensation You can use a saved compensation image whenever comp...

Page 35: ...hold settings Therefore the scanner will not use those areas for the creation of the compensation images Current threshold To modify the threshold settings set the desired threshold level in the box Use Selection Box Select this if you have a compensation slide that has only certain predefined areas for creating compensation images due to the type of the compensation slide or because the compensat...

Page 36: ...Pannoramic DESK II 1 23 User s Guide 4 Software Description Image Browser panel Figure 17 Image Browser This function displays the created compensation image June 4 2018 Rev 2 3DHISTECH Ltd 36 82 ...

Page 37: ...ation about the setting you are interested in Base Settings Figure 18 Base settings Objective type You can select the type and the magnification level of the objectives Camera type Select the correct camera type in Pos 1 Camera type then the software automatically loads the camera settings upon clicking the button June 4 2018 Rev 2 3DHISTECH Ltd 37 82 ...

Page 38: ...tandard Gamma value White Point D50 Legacy a Color scheme calibrated for the previous software version is applied Custom parameters are freely customizable as follows Color Scheme Calibrated Scheme is defined by the color calibration slide Legacy a Color scheme calibrated for the previous software version is applied Raw the color information of the recorded raw image is applied White Point D65 col...

Page 39: ...ESK II 1 23 User s Guide 4 Software Description Filter Settings Figure 19 Filter settings The settings on this panel apply to Fluorescent mode therefore irrelevant and inactive June 4 2018 Rev 2 3DHISTECH Ltd 39 82 ...

Page 40: ...amera rotation Caution If you have transported the product or changed the camera there is a possibility that the camera does not display the images properly If you experience such problems contact service support Do not modify these settings yourself June 4 2018 Rev 2 3DHISTECH Ltd 40 82 ...

Page 41: ...plication call settings for the following cases Slide Scanning Finished Scanning Error Occurred For Slide Scanning Finished and Scanning Error Occurred you can define the external software by entering the file path in the Application box or by browsing for it by clicking If you define parameters in the Parameters box it will pass them to the software that it starts If no parameter is defined then ...

Page 42: ...erver add or remove users and also set the message body in the Message subject field by entering a text for example to inform recipients on the location of the scanner or the slide set you are working with Click Test to test settings by sending an e mail to the address defined in the E mail To field after you have finished to check their relevancy before applying them First click Setup to display ...

Page 43: ...s button to send test e mail to the specified address There are 2 feedback messages that come up after sending the e mail An e mail has been sent to the recipient An error has been occurred The e mail could not be sent Check parameters then retry If you have properly set messenger settings click OK then click Users in the Message Call section of the Event Call Settings window to address functions ...

Page 44: ...y settings You can also set profile settings here and start scanning in Brightfield Manual mode For more information on profiles see section 4 5 Using Profiles Figure 24 Main window of control software 1 Tab panel with scanner functions and settings 2 Scan camera live image display area 3 Preview image recorded by the preview camera 4 Navigation and Preview image editor toolbar 5 Information area ...

Page 45: ...tions detailed in section 3 2 Caution Ensure that the slide is not tilted and that it rests perfectly horizontally on the stage To enter the slide name To enter the slide name manually type it into the Slide Name field To use the barcode as a slide name if a compatible barcode sticker is affixed on the slide click First the software automatically detects the barcode on the slide then the barcode s...

Page 46: ...n the size native resolution You can also find the button on this tab that you can start scanning with but it is recommended to go through all the settings before clicking it To administer profiles use the Scan Profiles field The current profile name is displayed at the bottom left corner of the screen To save the current profile click Save To select an existing profile from the selection menu cli...

Page 47: ... You can choose the following image formats in the Image Encoding field JPEG The quality can be adjusted with the JPEG quality factor slider Greater values result in better quality but larger files The default setting is 80 PNG lossless compressed image format BMP lossless uncompressed image format but larger files due to less effective compression It is a raw data file June 4 2018 Rev 2 3DHISTECH...

Page 48: ...erence images for all slides Use this function when the slide is completely covered by the specimen for example blood smear thereby making compensation image capture impossible For such slides use an empty slide with coverslip to grab a compensation image for more information see section 4 2 2 Image Compensation If you select this function the software will not create new compensation images for e...

Page 49: ...imen with auto threshold This automatically detects the specimen in the preview image Important Make sure that the detection result is acceptable for you Otherwise set the threshold value manually or append missing areas with the help of Scanned function located at the preview toolbar For more information see section 4 3 4 Preview area scanned by preview camera Preview Toolbar June 4 2018 Rev 2 3D...

Page 50: ...mple weak contrast it can help to encircle completely the specimen with a black marker pen You have to draw a full circle with the pen to enable the detection Marker pen detection will make specimen detection easier by limiting the scanned area If this option is selected only the area encircled with a marker pen will be scanned You can define the threshold level by changing the threshold value in ...

Page 51: ...reas Warning Do NOT use Remove coverslip function on TMA slides The layout of TMA slides may mislead the detecting algorithm Fill holes in scan area If selected the holes inside the specimen will also be scanned Often these holes contain small sample objects that otherwise would not be scanned Remove specks If selected the scanner does not digitize objects smaller than the predetermined size This ...

Page 52: ...or TMA grid Use selection box If selected only the area inside the box is scanned It is useful if you only want to scan the area that is equivalent with the size of the coverslip so external specks will not be scanned 4 3 4 Preview area scanned by preview camera The preview image shows the area that will be digitized with a red and or an orange mask If the Use selection box is enabled there is a b...

Page 53: ...will not be scanned These areas are not digitized on the basis of their position for example they are outside the selection box They are detected by the selected threshold value Marker If Scan area inside the marker is selected in the Preview tab the preview camera detects the marker pen The inside areas appear in red color and will be digitized Areas scanned in This color is only displayed during...

Page 54: ...pecimen holder that holds the current slide click Position sample and then click on the area of interest on the preview image This will define the Live image that you can access from the Focus panel To display the original image created by the preview camera click To shrink or enlarge the preview image set the zoom level at the left Size selection box 100 600 To fit the preview image to the previe...

Page 55: ... Enter the number of rows and columns in the Rows and Columns fields or select the appropriate values with the arrows 3 Enter the distance between two spots on the X and the Y axis in µm in the Spot distance X µm and Spot distance Y µm fields or select the appropriate values with the arrows 4 Enter the average diameter of the spots in the Diameter of spots µm field or select the appropriate value ...

Page 56: ...w image To move the spots enter the pixel value in the Move field or use the arrows To rotate the spots enter the degree value in the Rotate field or use the rotation arrows Figure 37 TMA tool Move tab with preview Label area To rotate the label area by 180 select the Rotate image by 180 degrees checkbox Figure 38 Label area June 4 2018 Rev 2 3DHISTECH Ltd 56 82 ...

Page 57: ...in an area that remained blurry with Auto Focus for example the specimen is thicker in that area The manual focus is mainly for setup purposes not for using the product as a traditional microscope Define the manual focus step size by adjusting the Step size slider Search for the sharpest focus on the current field of view automatically by clicking Multilayer mode Select Single layer option if you ...

Page 58: ...rvice tab Figure 41 Service tab This tab contains control and maintenance functions Show images during scanning If you select this option you will be able to see the currently scanned image field during scanning in the live image display field This slows down digitization process and should be activated only with purpose 4 3 7 Display field and toolbar for live images The display field for live im...

Page 59: ...avigation on the live image Click a desired area of the slide in the Live image to display that area in the center of the Live image field Toolbar for live images The toolbar is only available for the displayed live image but not available for camera images during digitization June 4 2018 Rev 2 3DHISTECH Ltd 59 82 ...

Page 60: ...Fast live mode live images are displayed with a frequency of at least 10 images per second If an exposure time is set to more than 100 ms the pixel values are multiplied accordingly for example if an exposure time of 200 ms is set the pixel values are multiplied by 2 and the exposure time is reset to 100 ms Therefore the resulting images will have a higher noise level Normal live mode images are t...

Page 61: ...errupt a running digitization process by clicking This button is only enabled if a digitization process is in progress 4 4 Scan To Server The ScanToServer function is available for you to scan directly to a dedicated CaseCenter server To setup connection with a designated server follow the steps below 1 Click Scan Options on the Routine Work tab panel Figure 43 Scan Options button 2 Select the Sca...

Page 62: ...enter option 3 Type in or select the server address from the Connect to drop down list and give the User name and Password Activate the Secure connection option to enable HTTPS connection Click Connect when finished Figure 45 Server Log in window June 4 2018 Rev 2 3DHISTECH Ltd 62 82 ...

Page 63: ...o close the Scan Options window Slide settings in the CaseCenter window 1 In the Routine work tab panel click the button 2 The CaseCenter window opens Figure 47 CaseCenter window Slide data is displayed in a form Upon activating the Read barcode if no name is defined on the list option the slide barcode will be read and set in the Barcode field If you do not want to use this option you can type a ...

Page 64: ...s of the barcode will be displayed and set as slide name 123 Case and Block fields can be set this way as well if necessary NOTE When filling slide data fields press TAB to move to the next cell Figure 48 Profile selection in the Scan To Server window Profile can be added by selecting a list entry from the Profile drop down list Select a folder where the slide will be scanned to After clicking the...

Page 65: ...atically selected as scanning destination based on the CaseID containing slide and patient data stored on HIS LIS Therefore Slide name Folder Case and Block field data cannot be modified only a profile can be selected Figure 50 CaseCenter window during Barcode parsing Click Done when finished or Cancel to quit the Scan To Server window If you save the form slides will be scanned to server accordin...

Page 66: ...size ON OFF available if Point Grey Grasshopper 3 camera is used Determination of the area to be scanned Use of automatic threshold specimen Use of specimen threshold value Use of marker pen threshold and specimen threshold values Remove coverslip function ON OFF Use of scan everything within marked area function ON OFF Filling of holes ON OFF Use of removing specks according to the set value ON O...

Page 67: ... currently valid legal requirements Dust specks and humidity may decrease or harm the operating capability of the device Protect it with dust cover when not in use Switch the power off before covering it Avoid touching the lamp bulb Wait for about 15 minutes to allow the bulb to cool down and disconnect the power plug before changing the bulbs Defective devices do not classify as domestic waste Di...

Page 68: ...ol software has stopped Name of currently selected slide for digitization already exists and is open in Pannoramic Viewer Close virtual slide in Pannoramic Viewer Slide jammed 1 Exit the application software 2 Switch off the power supply and disconnect the power cable 3 Remove the protective cover of the Pannoramic DESK II Follow the steps in section 5 3 1 Removing the protective cover 4 Remove sl...

Page 69: ...sitioned incorrectly Guiding fault in transport system Contact Thermo Fisher Scientific Service for support Digitization process is extremely slow Extended focusing option is activated Deactivate the Extended focusing option if not required Z stack option is activated Deactivate the Z stack option if not required Image is out of focus Focus range limit is activated Deactivate the Limiting focus ra...

Page 70: ...orm maintenance work and because of slide jamming the protective cover of the Pannoramic DESK II need to be removed in order to eliminate the problem First make sure that the running digitization procedure is interrupted prior opening the cover Before removing the cover exit the control software switch off the 24 V power supply and disconnect the power cables Caution Only instructed persons or ser...

Page 71: ...art of the cover must be pushed over the top part of the blue basic unit 7 5 To lock the cover in place insert the screws again and fasten them Caution Make sure that no cables are squeezed as you replace the cover s 5 3 2 Removing a slide If a slide got jammed or has suffered breakage during the process of digitization stop the process and if possible move the slide holder to a position where the...

Page 72: ... Warning If the slide is broken make sure to remove all pieces completely from the device The edge of the broken area of the slide is sharp and it can lead to injury and contagion Broken slide can be removed only by a well trained person 6 Swing fixing lever 3 back to the right towards the holder 7 Reattach the cover see section 5 3 1 Removing the protective cover 8 Connect the power cable of the ...

Page 73: ... persons can carry the base unit by holding it by its base plate You can find the weight data of the base unit in section 6 Technical Data Caution Be sure that the scan camera is not moved during transportation To transport Pannoramic DESK II do the following 1 Turn power supply off and detach all cables 2 Transport Pannoramic DESK II 3 After finishing transportation reestablish cable connections ...

Page 74: ... supply approx 550 g Ambient conditions Transportation shipment packed Permissible ambient temperature 40 C to 70 C Storage Permissible ambient temperature 0 C to 60 C Permissible relative air humidity non condensing 75 Operation Permissible ambient temperature 10 C to 35 C Permissible relative air humidity 75 Altitude at installation site max 2000 m Atmospheric pressure 500 hPa to 1060 hPa Degree...

Page 75: ...ete device 60 VA SINPRO MPU101 108 24V external mains power supply unit Intended site closed room facilities Electrical protection class I Internal protection degree IP 20 Electrical safety in compliance with UL 2601 1 TÜV EN 60601 CE EN60601 1 2 2001 Over voltage category II Radio interference suppression in accordance with CISPR 11 EN 55011 class B Line input voltage 100 V AC to 240 V AC Line in...

Page 76: ...tion with 20x objective and 0 63x C mount adapter 0 27µm Pixel resolution with 20x objective and 1x C mount adapter 0 17µm Pixel resolution with 40x objective and 0 63x C mount adapter 0 14µm Pixel resolution with 40x objective and 1x C mount adapter 0 09µm Connection type USB 3 0 Control computer minimum system requirements Operating System Microsoft Windows 7 SP1 64 bit EN CPU Intel Core i5 4590...

Page 77: ...he cables of the scanner 4 Remove the cover from the device by losing 1 1 screw on the side of the scanner 5 Remove the main camera and camera adapter 6 Place back and mount the cover of the scanner 7 Turn off and disconnect all the cables of the workstation and display Main parts and their accessories 1 BOX 1 Big carton box a fixing foams for the angles of the scanner box 2 BOX 2 DESK II scanner ...

Page 78: ...essories 3 pcs a dongles MATROX 3DView license driver CDs b main camera c cables d air duster sensor cleaning pen if any e objective cleaning papers if any f demonstration samples if any Packaging The whole system is packed in a big carton box This is divided into two areas with a vertical carton slab Space 1 for accessories PC cables etc Space 2 for BOX 2 scanner Figure 2 June 4 2018 Rev 2 3DHIST...

Page 79: ...ions 1 BOX 2 scanner box 2 Fixing foams for underneath BOX 2 3 Fixing foams for above BOX 2 4 Bottom part of BOX 1 5 Top of BOX 1 6 BOX 3 7 BOX 4 Figure 3 June 4 2018 Rev 2 3DHISTECH Ltd 79 82 For research and education uses only not for use in diagnostic procedures ...

Page 80: ...lding the scanner at its motherboard between its feet and taken inside the BOX 2 s lower part onto the foams inside it The shape of the rubber foam defines the orientation how they should put on After the scanner and foams are taken in you can place the upper part of the box of the scanner Now place the BOX 2 into BOX 1 and place the fixing rubber foams around Figure 4 Packing the accessories Plac...

Page 81: ...articularly large or bulky packages cannot be stored in warehouses or sheds they must be carefully covered with tarpaulins Recycling This indicates that an object is capable of being recycled CE The symbol shows that the device complies with the essential requirements of the relevant European health safety and environmental protection legislation in practice by many of the so called Product Direct...

Page 82: ...67 69 70 M Marker 49 52 67 P Pannoramic 6 Pannoramic DESK II 6 8 9 10 12 13 15 16 17 18 22 26 28 29 69 71 74 Pannoramic Viewer 17 58 69 70 Preview 19 21 27 29 34 43 48 50 51 52 53 55 56 58 77 R RGB 61 S Scan 12 14 17 34 43 45 46 47 48 49 50 51 52 58 59 60 62 67 70 74 77 Scan To CaseCenter 47 Scan To Server 65 66 Scanner 6 34 50 ScanToServer 47 62 Z Z stack 17 58 67 70 June 4 2018 Rev 2 3DHISTECH L...