Pannoramic 1000 1.2 - User’s Guide
8 Packaging instructions for Pannoramic
1000
8.3 Packaging
The scanner is packed in a huge wooden box. The computer and monitor have separate boxes (as
seen above).
Detailed images of the wood box can be found at the end of this document.
8.3.1 Packing the P1000 scanner
The most sensitive and most heavy unit is the P1000 scanner. Please handle it with care, the weight
of the device is more than 250 kg.
The structure of the BOX 1 including the foams is shown on
Figures 48-51
, and
55-58
. The fixing
points on the bottom of the box and the circles marked on the base for the instrument’s feet (see
Figure 51
) defines the orientation how the scanner should be placed inside, so first please examine it!
Where fixing points are closer to the edge is the front side. Front side is signed too with text.
Figure 48 – BOX 1
The scanner should be lifted up by holding the machine with a forklift or a mini-stacker at its bottom
between its feet and taken inside the BOX 1’s lower part onto the beams. Please check the position of
the instrument.
On the right there should be about 50 mm space from the edge of the bottom wood plate on the
base. (You can take Side B to its place on the right for a try – the foams should fill the ~50 mm gap –
but after checking the position you should remove Side B because of the steps below.)
Put rocker panels (if any) to their place into the foams behind the scanner. First slide the long front
panel then put the foam to the corner (
Figure 49
) then place the left and right panels into the foam
on the other side. (
Figure 50
)