PANNORAMIC® DESK FLASH DX
OPERATION
INFORMATION FOR USE
–
FOR INVESTIGATIONAL USE ONLY.
THE PERFORMANCE CHARACTERISTICS OF THIS PRODUCT HAVE NOT BEEN ESTABLISHED.
55
Inserting a glass slide in the Scanner
Figure 28: Inserting a magazine into the Scanner
1.
Click
to move the glass slide holder to loading position.
2.
Open the slide insert door (
1
) by sliding it to the left.
Caution!
When sliding the door, fingers may be pinched during operation.
3.
Open fixing lever by pulling it to the left.
4.
Insert slide (
3
) into the slide holder (
2
) by pinching it firmly with your thumb and
index finger at the label area (coverslip and the label area should be on the top).
Warning!
ALWAYS
be cautious when handling (removing or loading) a slide and avoid contact with
areas marked with
and
at the slide insertion area of the Scanner.
Always make sure that the slide is correctly inserted on the specimen holder. Incorrect slide
insertion may break the glass. The edge of the broken area of the slide is sharp and its
removal can lead to injury and contagion! Broken slide can be removed only with special
care!
5.
Once you have finished with loading the slide, Close the fixing lever by pushing it
to the right so that it touches and therefore fixes the slide
6.
Close the door by sliding it to the right.