PANNORAMIC® DESK FLASH DX
OPERATION
INFORMATION FOR USE
–
FOR INVESTIGATIONAL USE ONLY.
THE PERFORMANCE CHARACTERISTICS OF THIS PRODUCT HAVE NOT BEEN ESTABLISHED.
57
3.
Open the slide insert door (
Figure 28
–
1
) by pushing it to the left.
4.
Open the fixing lever by pulling it to the left.
5.
Remove the slide (
Figure 28
–
3
).
6.
Insert another slide or close the fixing lever and the door (
Figure 28
–
1
).
Checking the Scan Results
Warning!
It is
ALWAYS
the user’s responsibility to va
lidate scanning results by comparing the digital
slide with the macro image of the histological section or the tissue/sample on the glass slide.
Note, that there can be differences between the digital image of the slide and the macro
image.
Follow these steps to view the digitized image of the slide:
1.
Select the slide in Slide View first, then double-click the thumbnail image at the left of the
slide information bar to open the digital slide in the Viewer program.
2.
Make sure that the digital slide includes all the tissue area selected for scanning.
Checking the Status of the Scanner
Slide statuses are indicated on the Slide View of the
PANNORAMIC® Diagnostic Scanner
Software interface.
Notice:
Slide status is presented real-time in the Slide View of the P
ANNORAMIC® Diagnostic
Scanner Software. If an error has occurred during the scanning process, always make sure to
read information displayed on the interface before removing the slide that is involved in a
faulty or failure event.