PANNORAMIC® DESK FLASH DX
TECHNICAL DATA
68
INFORMATION FOR USE
–
FOR INVESTIGATIONAL USE ONLY.
THE PERFORMANCE CHARACTERISTICS OF THIS PRODUCT HAVE NOT BEEN ESTABLISHED.
Pannoramic® DESK FLASH DX
General Performance
Throughput
–
Scanning time of a 15
×15mm benchmark tissue size at
20
×
equivalence is ~35
seconds
, and at 40× equivalence is ~120 seconds (15×1
5 mm area, with adaptive focus point
algorithm, local scanning, single layer, JPEG compression, MRXS file format)
Tissue detection
–
AI-based
Glass Slides
Definition
Slide
Coverslip
Length
75.0 to 76.0 mm
max. 50 mm
Width (standard slide)
25.0 to 26.0 mm
max. 24 mm
(recommended: 22 mm)
Width (large slide)
51.0 to 52.0 mm
max. 50 mm
Thickness
0.90 to 1.2 mm
No. 1 and No. 1.5 (0.13 to 0.16 mm and 0.16 to 0.19 mm)
Corners
45° beveled corners
N/A
Edges
Ground or cut
N/A
Table 7: Glass slide specifications
Digital Slide Images
Definition
Specification
Pixel size (using
20× objective and
1.0
× C
-mount adapter)
0.24
µm
Pixel size (using 4
0× objective and
1.0
× C
-mount adapter)
0.12
µm
Default image compression level
80%
Output formats
Lossless/RAW, DICOM, MRXS
Label image format
JPG
Table 8: Digital slide images specifications