PANNORAMIC® DESK FLASH DX
TECHNICAL DATA
INFORMATION FOR USE
–
FOR INVESTIGATIONAL USE ONLY.
THE PERFORMANCE CHARACTERISTICS OF THIS PRODUCT HAVE NOT BEEN ESTABLISHED.
69
External Power Supply Unit
Definition
Specification
Mains type
Single phase load (hot, service neutral, separate ground)
Input voltage range
100-240 Vac
Line input frequency
47/63 Hz
Max freq. deviation
±1 Hz
Secondary voltage
24V DC
Secondary current
max. 4.8 A
Table 9: Power supply specifications
Notice:
Wall outlets must have a ground connection. The power mains must be hospital grade in the
United States and Canada. In other countries, the connector must be approved by the
relevant local safety regulations for this application.
Electromagnetic Compatibility
The Scanner is intended for use in the electromagnetic environment specified below. The owner
of the Scanner must assure that it is used in such environment.
RF emissions - Group1
–
The Scanner RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment. Class B
Network and other Connections
The Scanner / control computer supports the following connections:
•
Connection to network/server
–
10 Gb/s Ethernet cable class CAT6A with RJ45
connector
•
Control
–
USB cable (between Scanner and control computer)
•
Preview camera
–
USB cable (between Scanner and control computer)
•
Adimec
–
CoaXPress cables (between Adimec camera and control computer)