Pannoramic SCAN II 2.2.0 User’s Guide
6.2 Safety Information
Slides that are not used any longer do not classify as domestic waste and must be treated as
hazardous materials and disposed in compliance with the currently valid legal requirements.
Dust, specks and humidity may decrease or harm the operating capability of the device. Protect it
with dust cover when not in use. Switch the power off before covering it.
Avoid touching the lamp bulb. Wait for about 15 minutes to allow the bulb to cool down and
disconnect the power plug before changing the bulbs.
There is crush hazard during operation in the area of input or output. Inserts/magazines should be
moved through software. Never open the device unless prompted accordingly by the software.
Defective devices do not classify as domestic waste. Dispose of the device in compliance with the
currently valid legal requirements.
Lumencor SPECTRA 6 illumination unit
•
NEVER
switch on the Lumencor SPECTRA 6 Light Engine illumination unit when it is not
connected to the microscope.
•
Ensure that the light source is properly connected.
•
Do not cover the unit, as it would cause abnormal temperature rising.
Lumencor SOLA SM II illumination unit
•
Keep inflammable and combustible materials outside the range of
Lumencor SOLA SM II
illumination unit.
•
Do not expose yourself to the UV radiation of the
Lumencor SOLA SM II
illumination unit.
•
When the
Lumencor SOLA SM II
illumination unit is switched on:
◦
NEVER
look directly into the fiber-optic cable.
◦
NEVER
remove the fiber-optic cable from the device.
◦
NEVER
connect the fiber-optic cable to the device.
◦
NEVER
remove the housing from the device.
6.2.1 Cleaning the device
•
Before cleaning, switch off the device and pull the plug out.
•
Do not let liquid penetrate into the system.
•
To clean liquid from the device, use a lint-free cotton cloth. You can moisten the cloth with
water.
•
To clean greasy surfaces, it is recommended to apply
soapy water
on a piece of lint-free
cotton cloth and gently rub the surface.
January 24, 2020 - Rev. 2
3DHISTECH Ltd.
134(156)