6.3 Troubleshooting
7. Switch on the 24 V power supply.
8. Switch on the control computer.
9. Launch control software.
•
Magazine jammed
1. Exit control software.
2. Switch off control computer and 24V power supply, disconnect power cable.
3. Remove protective covers (see chapter
6.4.1 Removing the protective covers
).
4. Remove magazine manually (see chapter
6.4.4 Removing the Magazine from the
). If a jammed magazine cannot be removed, contact service
personnel.
5. Reattach protective covers.
6. Reconnect 24V power supply and control computer to line power supply.
7. Switch on the 24 V power supply.
8. Switch on the control computer.
9. Launch control software.
Virtual slide is improperly focused almost across its entire surface
•
Focus range enabled
•
Disable
Limiting focus range
.
Poor image quality
•
Imaging error (for example, chromatic aberrations)
•
Contact 3DHISTECH Service for support.
Warning!
Do not modify the camera driver settings, as the default factory
settings give the most accurate images.
Virtual slides show poor stitching in Pannoramic Viewer
•
Scan camera not properly adjusted (for example, due to loose fixture)
•
Contact 3DHISTECH Service for support.
January 24, 2020 - Rev. 2
3DHISTECH Ltd.
136(156)