Pannoramic SCAN II 2.2.0 User’s Guide
6.3 Troubleshooting
Slide is often positioned incorrectly
•
Guiding fault in transport system
•
Check if faulty positioning disappears after the selection of a different type of
magazine (for example, universal magazine).
•
Contact 3DHISTECH Service for support.
Digitization process is extremely slow
•
Extended focusing
option is activated
•
Deactivate the
Extended focusing
option if not required.
•
Z-stack
option is activated
•
Deactivate the
Z-stack
option if not required.
Image is out of focus (in Manual mode)
•
Focus range limit is activated
•
Deactivate the
Limiting focus range
option.
•
Focus range setting is incorrect
•
Change focus range setting.
All individual images show a shading effect
•
Algorithm failed to find empty image fields
•
Capture a reference image with a blank slide (see chapter
).
6.3.1 Fluorescent Applications (optional)
The switching of the Lumencor SPECTRA 6 cannot be triggered, in spite of trigger connection
through the camera.
•
Contact 3DHISTECH Service for support.
No suitable exposure time can be found. Signal is too weak.
•
Lumencor SOLA SM II is switched off
•
Switch light source on, otherwise contact 3DHISTECH Service for support.
January 24, 2020 - Rev. 2
3DHISTECH Ltd.
137(156)