Pannoramic SCAN II 2.2.0 User’s Guide
2 Installation
•
USB-to-RS232 serial communication cable (between Lumencor SPECTRA 6 fluorescent
illumination unit and control computer)
Warning!
Only connect external devices that are safety extra low voltage rated to
the instrument to avoid the risk of electrical shock.
2.4 Coupling the Lumencor SOLA SM II unit with the scanner
To operate the Lumencor SOLA SM II fluorescent illumination unit, it has to be connected to the
Pannoramic® SCAN II base unit.
1. Turn the control computer and the Pannoramic® SCAN II unit off.
2. Connect both ends of the fiber optic cable to the coupling ports of the scanner and the
illumination unit (see
Figure
2 – option A
).
3. Connect power supply cables both of the scanner and the illumination unit.
2.5 Coupling the Lumencor SPECTRA 6 unit with the scanner
To operate the Lumencor SPECTRA 6 Light Engine fluorescent illumination unit, it has to be connected
to the Pannoramic® SCAN II base unit, the camera, and the control computer.
1. Turn the control computer and the pco.edge 4.2 camera off.
2. Attach Lumencor SPECTRA 6 fluorescent illumination unit to the coupling adapter of the
Pannoramic® SCAN II scanner (see
Figure
2 – option B
).
3. Connect power supply cables both of the scanner and the illumination unit.
4. Connect TTL trigger cable between pco.edge 4.2 camera (out 4) or Point Grey Grasshopper 3
camera and the Lumencor SPECTRA 6 unit (TTL port), and USB-to-RS232 communication
cable between the Lumencor SPECTRA 6 unit and the computer.
Figure 5 – Lumencor SPECTRA 6 Light Engine connecting ports
January 24, 2020 - Rev. 2
3DHISTECH Ltd.
26
(