PANNORAMIC® SCAN RX 1.1
SETTINGS AND SERVICES
NOVEMBER 18, 2021
124
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.
For
Slide scanning has finished
and
Scanning error has occurred
you can define
the external software by entering the file path in the
Application
box, or by
browsing for it by clicking
.
If you define parameters in the
Parameters
field, it will pass them to the software
that it starts. If no parameter is defined, then the parameter will not be passed.
The default parameters are the following:
▪
For
Slide scanning has finished
: "<SLIDENAME>"
–
transfers the slide
name
▪
For
Scanning error has occurred
: "<ERRORTEXT>"
–
transfers the error
message
Notice:
Pannoramic® Scanner Software for Research control software can provide not only the
“<SLIDENAME>” but the path of the digital slide as well
.
Run
defines the window mode that the external application will start in. The
following window modes can be set:
•
Normal window
•
Minimized
•
Maximized
You can test whether the
Scanning Error Occurred
settings work by clicking
.
Caution!
If
C:\ProgramData\3DHISTECH\SlideScanner\MessengerUserPreferences.xml
is missing or
damaged, the
External application call
feature will not be able to read the address book,
therefore it will not operate. To solve this issue, the address book must be repaired or
created again.