PANNORAMIC® SCAN RX 1.1
SOFTWARE DESCRIPTION
NOVEMBER 18, 2021
37
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.
4.2
Main window functions
Figure 12. Pannoramic® Scanner Software for Research 1.1 main window
The following functions are available for you to select from at the main menu bar:
Switch between
Microscope
and
Planner
modes
Microscope initialization
–
All motors and the preview camera during start-up
Switch between
Brightfield mode
and
Fluorescence mode
–
The camera set for the
selected mode is initialized.
Slide/magazine control
Notice:
At slide loading, the Pannoramic® Scanner Software for Research control software
registers, whether a slide is inserted or not.
At the initialization, the stage moves to the slide insert position and checks the slide.
A logging mechanism is also available in the slide loading test.
The slide sensor has an optoelectronic gate, and a lever. At the slide loading position, the
lever is lifted into the gate and shutters the light beam. If there is no slide in the loader arm,
the lever cannot be lifted into the gate and this way, the light beam is not shuttered.
Upon clicking the slide icon or entering slide position number at the right side and clicking
Load
, the slide can be loaded.