Service Manual
Fire
CR
+
Computed Radiography Scanner
22
Parts Removal/Replacement |
Fire
CR
+
Optic Plate
1.
Remove Top Cover. (Pg. 14)
2.
Remove Internal Cover. (Pg. 16)
3.
Remove 2 screws from each of the three spring loaded brackets (2.5mm hex).
Note: Keep
pressure over the spring while removing and re-installing screws.
4.
Unplug the 3 wire connectors (1 to BLDC, 2 to Main Board)
5.
Lift out Optic Plate.
6.
Reinstall in reverse order.