| Troubleshooting
39
Mechanical Noise
Belt Jumping
Bent main board plate
1.
Remove Top Cover. (Pg. 14)
2.
Remove internal frame from base.
3.
Remove two spike fastener clips holding the main board plate.
4.
Remove 4 hex screws.
5.
Once screws and clips are removed from the main board plate begin to
push up towards the board until the plate is straight again.
6.
Reinstall in reverse order.
Belt tension
1.
Remove Top Cover. (Pg. 14)
2.
Remove the Internal Cover (Pg. 16) and locate the Y-axis motor.
3.
Loosen the two Allen screws on the Y-axis motor’s base.
4.
Pry the Y-axis motor forward keeping it straight until the tension on the
belt is fairly tight.
5.
Tighten the screws.
6.
Reinstall in reverse order.
Unaligned stage
1.
Remove Top Cover. (Pg. 14)
2.
Remove the Internal Cover. (Pg. 16)
3.
Loosen the two Allen head screws on each side of the main axle pulley.
4.
Move the stage assembly around.
5.
Slide the stage assembly all the way to the back of the scanner.
6.
Tighten the Allen screws on each pulley.
7.
Reinstall in reverse order.
Broken cassette lock
1.
Replace cassette.