EO Abator Operation and Installation Manual
20
Figure 4
To connect 3M Steri-Vac Models 5XL, 8XL, and GS Series Sterilizer(s) to EO Abator
If only one 5XL/8XL/GS Series sterilizer is connected, connect its wires as follows: A jumper wire is
connected from abator terminal 27 to 25 and from terminal 28 to 26.
If a second sterilizer is connected, connect wires to these terminals:
8XL / 5XL
“D” Connector
Sterilizer #2
PIN #
4XL / 5XL
3M Steri-Vac
Sterilizer #2
Wire #
to
EO Abator System
Indicator Panel
Terminal #
10
20
15
15
22
22
-
22
26
13
25
25
4
38
16
14
40
21
Note:
One 5XL/8XL/GS Series “D” connector cable (part number 78-8078-6209-5) is provided with
each EO Abator. Additional cables are available through the 3M Health Care Service Center.
Note:
Use shielded cable (Belden 18 AWG Part #83656).