13
Bair Hugger Model 750 Temperature Management Unit Service Manual
English 202522A
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document.
www.bairhugger.com reorder #202522A
Method
Note: If the temperature management unit has been stored at a temperature lower than 20°C, allow it to
reach ambient temperature before beginning the calibration procedure�
1� Attach the Model 22110 temperature test device to the end of the temperature management unit hose�
2� Connect the temperature management unit to a grounded power source� The unit performs a self-test
and automatically proceeds to
Standby
mode�
3� Press the Ambient button to place the unit in
Ambient
mode� Allow the unit to run in
Ambient
mode
for at least ten minutes to equalize the temperature of the unit’s internal components�
4� Press the Alternative Mode button (see Figure 2) and the Ambient button simultaneously for
3 seconds� The temperature management unit is now in
Calibration
mode� In this mode, the
temperature management unit performs the following tasks:
• 37
°C and
38
°C indicator lights flash�
• Blower activates�
• Heater elements deactivate�
• The text “TMP” and the temperature in °C (0�1°C resolution) at the end of the hose appear in the
alphanumeric display�
• The text “OS” and the current calibration offset value (0�1°C resolution) appear in the
alphanumeric display�
5� Compare the temperature shown on the alphanumeric display and the temperature on the
independent, temperature calibration device�
• If the temperatures match, press the
Alternative Mode
button to return the temperature
management unit to
Standby
mode�
• If the temperatures do not match, adjust the temperature management unit’s displayed
temperature:
a� Press the
32°C
button to decrease the temperature or the
38°C
button to increase the
temperature by 0�1°C� You can adjust the temperature of the temperature management unit a
maximum of ±3°C�
b� Continue to press the
32°C
or
38°C
button(s) until the temperature in the alphanumeric display
matches the temperature on the calibration device�
c� Press the
Alternative Mode
button to store the calibration offset value and return the
temperature management unit to
Standby
mode�
Note: If it is necessary to enter an offset value greater than ±3°C, there may be a defect in the hose end
sensor� Remove unit from service until a new hose assembly can be substituted�