17
Bair Hugger Model 750 Temperature Management Unit Service Manual
English 202522A
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document.
www.bairhugger.com reorder #202522A
Fault Code Table
The software will detect and report the following conditions by displaying the corresponding fault code:
Fault Condition ____________________________________ Fault Code
Internal sensor 1 over-temperature condition
001
Hose end sensor 2 over-temperature condition
002
Hose end sensor 3 over-temperature condition
003
Internal sensor 1 shorted
004
Hose end sensor 2 shorted
005
Hose end sensor 3 shorted
006
Hose end sensor 2 open
007
Low wattage heater failure
008
Medium wattage heater failure
009
High wattage heater failure
010
A/D converter timeout
011
A/D startup check failed
012
(Reserved Fault Code)
013
Error in execution loop
014
Timeout on alphanumeric display busy flag clearing
015
RAM corruption in duplicate RAM locations
016
Error in executing a depressed key
017
Stuck key during startup
050
EEPROM is at max # write cycles
051
Non-0 EEPROM byte detected after zeroing
052
EEPROM checksum failure on copy # 1
053
EEPROM compare failure copy #1 to copy #2
054
EEPROM compare failure copy #1 to RAM
055
EEPROM write cycle time out
056
SPI byte transfer time out
057
EEPROM write error when entering standby
059
Real time clock failure
060
Temperature out of range high
070
Temperature out of range low
071
5 V power supply in excess of 5�333 V
100
5 V power supply under 4�667 V
101
Startup test EEPROM error
102
Startup test RAM error
103