Bair Hugger Model 750 Temperature Management Unit Service Manual
18
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document.
www.bairhugger.com reorder #202522A
Viewing the Hour Meter
Service frequency
As often as necessary�
Tools/equipment
None required�
Method
1� If your unit is already ON, you can begin the test from any operating mode� Otherwise, connect
the temperature management unit to a grounded power source� The unit performs a self-test and
automatically proceeds to
Standby
mode� Then select any operating mode�
2� Press the
Alternative Mode
button for three seconds (see Figure 2)� The temperature management
unit is now in
Hour Meter Display
mode� In this mode, the temperature management unit performs
the following tasks:
a� Continues to operate in current operating mode�
b� “Hours” appears in the alphanumeric display�
c� Cumulative hours of temperature management unit operation appear in the alphanumeric display�
3� Wait 60 seconds or press the
Alternative Mode
button to return the temperature management unit to
its previous operating mode�
Viewing the Operating Timer
To view the amount of time the temperature management unit has been operating in the current
temperature mode, press and hold the
Current Temperature
mode button for at least three seconds� The time
is displayed in HH:MM:SS format�