19
Bair Hugger Model 750 Temperature Management Unit Service Manual
English 202522A
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document.
www.bairhugger.com reorder #202522A
Replacing the Air Filter
Service frequency
Every 12 months or 500 hours of use�
Tools/equipment
• Replacement air filter
• #2 Phillips screwdriver
warning
Do not attempt to clean the air filter as it may contain hazardous microorganisms� Discard the filter in a
manner consistent with institutional protocol for biohazardous material�
Method
1� Disconnect the temperature management unit from the grounded power source�
2� Turn the temperature management unit upside down�
3� Remove the 2 screws located in the louvered filter cover�
4� Lift the filter cover off the temperature management unit�
5� Remove the air filter and discard it�
6� Place the new filter in the filter compartment with the gasket toward the filter ledge�
7� Replace the filter cover with the louvers facing the back of the temperature management unit�
8� Replace the 2 screws in the filter cover�