Bair Hugger Model 750 Temperature Management Unit Service Manual
30
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document.
www.bairhugger.com reorder #202522A
Electrical Characteristics
heating element
1400 W Resistive
leakage current
Meets UL 60601-1 and IEC 60601-1 requirements�
blower motor
Operating speed: approximately 4,000 rpm
Airflow: up to 48 cfm or 23 L/s
power consumption
Peak: 1550 W
Average: 800 W
power cord
15 ft�, SJT, 3 cond�, 13 A
15 ft�, SJT, 3 cond�, 15 A
4�6 m, HAR, 3 cond�, 10 A
device ratings
110-120 VAC, 50/60 Hz, 11�7 A, or 220-240 VAC, 50/60
Hz, 7�2 A, or 100 VAC, 50/60 Hz, 15 A
fuses
Type:
Fast acting ceramic fuses, 250 VAC
Amp rating:
12 A (110 - 120 VAC)
8 A (220 - 240 VAC)
15 A (100 VAC)
Operating speed:
F (Fast Acting)
Breaking capacity:
15 A, 12 A: 750 A @ 250 Vac
15 A, 12 A: 10,000 A @ 125 Vac
8 A: 200 A @ 250 Vac
8 A: 10,000 A @ 125 Vac
certifications
IEC 60601-1; EN 60601-1-2; UL 60601-1;
CAN/CSA-C22�2, No� 601�1, EN 55011;
EN 80601-2-35
classification
MEDICAL — GENERAL MEDICAL EQUIPMENT
AS
TO ELECTRICAL SHOCK, FIRE AND MECHANICAL
HAZARDS ONLY IN ACCORDANCE WITH UL
60601-1; CAN/CSA-C22�2, No�601�1; ANSI/AAMI
ES60601-1:2005 CSA-C22�2 No� 60601-1:08; EN 80601-
2-35; Control No�4HZ8�
Classified under IEC 60601-1 Guidelines (and other
national versions of the Guidelines) as Class I, Type BF,
Ordinary equipment, Continuous operation� Not suitable
for use in the presence of flammable anesthetic mixtures
with air or with oxygen or nitrous oxide� Classified by
Underwriters Laboratories Inc� with respect to electric
shock, fire and mechanical hazards only, in accordance
with UL 60601-1, EN 80601-2-35 and in accordance with
Canadian/CSA C22�2, No� 601�1� Classified under the
Medical Device Directive as a Class IIb device�
diagnostics
A qualified service technician can perform over-
temperature detection system testing, temperature output
testing, operating temperature calibration, and fault code
troubleshooting�