31
Bair Hugger Model 750 Temperature Management Unit Service Manual
English 202522A
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document.
www.bairhugger.com reorder #202522A
Maintenance Checklist
Filter inspection
1.
Remove and inspect the filter.
2�
Return filter or replace as necessary�
Note: Replace every 12 months or every 500 hours (PN 90047).
Calibrate the hose end sensor
1�
Attach the Bair Hugger temperature test unit (PN 22110)�
a�
Connect temperature sensor to your meter�
2�
Run in
Ambient
mode for 10 minutes�
3�
Enter the
Calibration
mode�
a�
Press the Alt (Bair Hugger logo) and Ambient button, together for
approximately 3-seconds�
b�
Temperature measured at end of hose will display on LCD�
4�
Adjust Bair Hugger unit's displayed temperature so it matches test unit sensor
(± �1° increments)�
a�
32
°
C button lowers displayed temperature�
b�
43
°
C button raises displayed temperature�
Note: Unit can only adjust ± 3.0°C. Try a different hose if more adjustment is
required and call 3M Patient Warming technical support with results.
5�
Save Settings – Press the Alt (Bair Hugger logo) button to save your settings�
Test over temperature detection system
1�
Initiate over-temperature test
a�
Press the Alt (Bair Hugger logo) and 43°C button together for approximately
3-seconds�
b�
Wait as unit executes self test�
c�
Two beeps indicate test has completed�
2�
Review Results
a�
Observe max temperature <56°C�
b�
Pass or fail indicated on display�