background image

12

3M™ Bair Hugger™ Warming Unit Model 775 – Service Manual

C

A

UTIONAR

Y

ERRORS

FA

UL

T

CONDITION

W

H

IL

E U

N

IT

 IS I

N U

SE

FAULT CODE

POSSIBLE CAUSE

SOLUTION

FC-001

FC-003
FC-008

Airflow occluded.

Disconnect from power source. Remove warming unit hose occlusion. Position

hose end correctly into hose card (refer to the Operator’s Manual). Check for

air intake obstructions on the bottom of the warming unit. Replace filter. Wait

30-60 seconds before reconnecting to the main power source. If no fault occurs,

warming unit is ready for use.

Intermittent elbow sensor

failure.

Disconnect from power source. Allow warming unit to cool. Wait 30-60 seconds

before reconnecting to the main power source.
If these faults continue to occur during use and the airflow is not occluded

replace elbow sensor.

FC-002
FC-003
FC-004
FC-005
FC-006
FC-007

Warming unit hose end

sensor failure at patient end.
Moisture.
Intermittent hose sensor

failure.
High heat.

Disconnect from power source. Allow warming unit to cool. Wait 30-60 seconds

before reconnecting to the main power source. If no fault occurs, warming unit is

ready to use.
If these faults continue to occur during use, replace hose.

Intermittent elbow sensor

failure.

Disconnect from power source. Allow warming unit to cool. Wait 30-60 seconds

before reconnecting to the main power source.
If these faults continue to occur during use and the airflow is not occluded

replace elbow sensor.

FC-070

Warming unit hose end

temperature exceeds the

set point by at least 5°C for

approximately 5 minutes.
Faulty control board.

Unplug the warming unit, then wait 30-60 seconds before reconnecting to the

main power source. If no fault occurs, warming unit Is ready to use.
Replace control board.
Calibrate and return to service.

FC-071

Warming unit hose end

temperature is at least 10°C

below the set point for

approximately 10 minutes.
Faulty control board.
Faulty heater.
Faulty blower.

Unplug the warming unit, then wait 30-60 seconds before reconnecting to the

main power source. If no fault occurs, warming unit Is ready to use.
Replace control board.
Replace heater.
Replace blower.
Calibrate and return to service.

Whistling sound during

operation

Air leaks in enclosure or

warming unit hose.
Faulty bearing in

blower motor.

Tighten all screws.
Check for holes in warming unit hose.
Check for loose warming unit hose collar and presence of O-rings.
Clean the filter gasket and mating surface.

 

 

Technical Support and Order Placement

USA

TEL:

800-733-7775

OUTSIDE OF THE USA

Contact your local 3M representative.

When You Call for Technical Support

We will need to know the serial number of your Bair Hugger warming unit

when you call us. The serial number label is located on the back of the

warming unit.

In‑Warranty Repair and Exchange
USA

Call 3M technical support if your warming unit requires factory service. A

technical support representative will give you a Return Authorization (RA)

number. Please use this RA number on all correspondence concerning

your warming unit. Your technical support representative will also send a

shipping carton to you at no charge, if needed. Call your local supplier or

sales representative to inquire about borrowing a warming unit while we

service your warming unit.

OUTSIDE THE USA

Contact your local distributor concerning in-warranty repair and exchange.

Returning Warming Units for Service
TOOLS/EQUIPMENT

3M service carton
Tape dispenser
Shipping label
Marking pen

METHOD

1.

Call 3M technical support (see page 12 for contact information) to get

a Return Authorization (RA) number and a service carton.

2.

Remove the top foam piece. If the plastic liner is available, use it to

enclose the warming unit.

3.

Lower the warming unit carefully into the carton.

4.

Replace the top foam piece. Observe proper orientation. Hose will

extend out in the corner of foam.

5.

Lay the warming unit hose on top of the foam piece around the edge.

6.

Seal the carton with tape.

7.

Apply the shipping label addressed to 3M Health Care.

8.

Write the RA number on the outside of the carton.

9.

Ship to 3M Health Care.

13

GB / 34-8718-8817-7

Specifications

Guidance and manufacturer’s declaration – electromagnetic emissions

The Model 775 warming unit is intended for use in the electromagnetic environment specified below. The customer or the user of the Model

775 warming unit should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment ‑ guidance

RF emissions
CISPR 11

Group 1

The Model 775 warming unit uses RF energy only for its internal function. Therefore, its RF emissions are

very low and are not likely to cause any interference in nearby electronic equipment.

RF emissions
CISPR 11

Class B

The Model 775 warming unit is suitable for use in all establishments, including domestic establishments

and those directly connected to the public low-voltage power supply network that supplies buildings

used for domestic purposes.

Harmonic emissions
IEC 61000-3-2

Class A

Voltage fluctuations/

Flicker emissions
IEC 61000-3-3

Complies

Guidance and manufacturer’s declaration – electromagnetic immunity

The Model 775 warming unit is intended for use in the electromagnetic environment specified below. The customer or the user of the Model

775 warming unit should assure that it is used in such an environment.

Immunity Test

IEC 60601 test level

Compliance level

Electromagnetic environment ‑ guidance

Electrostatic

discharge (ESD)
IEC 61000-4-2

±6 kV contact
±8 kV air

±6 kV contact
±8 kV air

Floors should be wood, concrete or ceramic tile. If floors are covered with

synthetic material, the relative humidity should be at least 30%.

Electrical

fast transient/burst
IEC 61000-4-4

±2 kV for power

supply lines
±1 kV for input/output

lines

±2 kV for power

supply lines
±1 kV for input/output

lines

Mains power quality should be that of a typical commercial or

hospital environment.

Surge
IEC 61000-4-5

±1 kV line to line
±2 kV line(s) to earth

±1 kV line to line
±2 kV line(s) to earth

Mains power quality should be that of a typical commercial or

hospital environment.

Voltage dips, short

interruptions and

voltage variations

on power supply

input lines
IEC 61000-4-11

<5%

U

T

(>95% dip in

U

T

)

for 0,5 cycle
40%

U

T

(60% dip in

U

T

)

for 5 cycles
70%

U

T

(30% dip in

U

T

)

for 25 cycles
<5%

U

T

(>95% dip in

U

T

)

for 5 sec

<5%

U

T

(>95% dip in

U

T

)

for 0,5 cycle
40%

U

T

(60% dip in

U

T

)

for 5 cycles
70%

U

T

(30% dip in

U

T

)

for 25 cycles
<5%

U

T

(>95% dip in

U

T

)

for 5 sec

Mains power quality should be that of a typical commercial or hospital

environment. If the user of the Model 775 warming unit requires continued

operation during power mains interruptions, it is recommended that the

Model 775 warming unit be powered from an uninterruptible power supply

or a battery.

Power frequency

(50/60 Hz)

magnetic field
IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should be at levels characteristic of a

typical location in a typical commercial or hospital environment.

NOTE

U

T

is the a.c. mains voltage prior to application of the test level.

PDF Scaled at 75.0%

Summary of Contents for Bair Hugger 775

Page 1: ... license in Canada 2016 3M All rights reserved Issue Date 2016 08 34 8718 8817 7 3M Deutschland GmbH Health Care Business Carl Schurz Str 1 41453 Neuss Germany Model 775 Warming Unit Service Manual Normothermia System 1 3M Bair Hugger Warming Unit Model 775 2 Appareil de réchauffement Bair Hugger 3M modèle 775 3 3M Bair Hugger Wärmeeinheit Modell 775 PDF Scaled at 75 0 ...

Page 2: ...3MTM Bair Hugger Patient Adjustable Warming Unit Requester Frank Boeshart Debra Staack Reference Die Doc Size 8 5 x 11 Item Spec 34 8718 8817 7 Supersedes Cat Product Model 775 Structure Service Manual PA 34871888177 indb GA InDesign CC 2014 BOOK PROCESS BLACK PMS 280 4394547 3 08 04 16 New Build kmh 08 26 16 Changes per RA proof report TN 09 15 16 Final release DJ PDF Scaled at 75 0 ...

Page 3: ...ernative Modes 6 Accessing the Alternative Modes 7 Exiting an Alternative Mode 7 Required Preventative Maintenance 7 Calibrating the Operating Temperature 7 Testing the Over Temperature Detection System 8 Viewing the Fault Code Log 8 Clearing the Fault Code Log 8 Fault Code Table 9 Viewing the Hour Meter 9 Viewing the Operating Timer 9 Replacing the Air Filter 9 Replacing the Hose 9 Detaching Warm...

Page 4: ... with the warming unit refer to the respective Instructions for Use Indications The Bair Hugger temperature management system is intended to prevent and treat hypothermia In addition the temperature management system can be used to provide patient thermal comfort when conditions exist that may cause patients to feel too warm or too cold The temperature management system can be used with adult and ...

Page 5: ... with the warming unit refer to the respective Instructions for Use Indications The Bair Hugger temperature management system is intended to prevent and treat hypothermia In addition the temperature management system can be used to provide patient thermal comfort when conditions exist that may cause patients to feel too warm or too cold The temperature management system can be used with adult and ...

Page 6: ...ing blanket gown The Bair Hugger warming unit is intended to be used with the hose properly attached to its corresponding warming blanket gown and in accordance with good practices for operating room sterile technique CAUTION To reduce the risk of patient or caregiver injury If mounted on an IV pole the distance from the bottom of the warming unit to the floor must be less than 44 112 cm and the I...

Page 7: ...ing blanket gown The Bair Hugger warming unit is intended to be used with the hose properly attached to its corresponding warming blanket gown and in accordance with good practices for operating room sterile technique CAUTION To reduce the risk of patient or caregiver injury If mounted on an IV pole the distance from the bottom of the warming unit to the floor must be less than 44 112 cm and the I...

Page 8: ...ation mode allows the technician to view important system information including the firmware system identification serial port baud rate 5V power supply status and the control board serial number The Over temperature Detection Test mode allows the user to test the operation of the independent over temperature shutdown system Accessing the Alternative Modes To access the alternative modes you must ...

Page 9: ...ation mode allows the technician to view important system information including the firmware system identification serial port baud rate 5V power supply status and the control board serial number The Over temperature Detection Test mode allows the user to test the operation of the independent over temperature shutdown system Accessing the Alternative Modes To access the alternative modes you must ...

Page 10: ...urs and minutes 3 Press either the 32 C or 38 C button to view the next fault code in the log The warming unit stores a maximum of 5 fault codes in non volatile memory 4 Press the Alternative Mode button to return the warming unit to its previous operating mode Clearing the Fault Code Log Put the warming unit in Fault Code mode by pressing the Alternative Mode button and the 32 C button simultaneo...

Page 11: ...urs and minutes 3 Press either the 32 C or 38 C button to view the next fault code in the log The warming unit stores a maximum of 5 fault codes in non volatile memory 4 Press the Alternative Mode button to return the warming unit to its previous operating mode Clearing the Fault Code Log Put the warming unit in Fault Code mode by pressing the Alternative Mode button and the 32 C button simultaneo...

Page 12: ...ordance with hospital practices for cleaning OR equipment After every use wipe the warming unit the outside of the warming unit hose and any other surfaces that may have been touched Use a damp soft cloth and a hospital approved mild detergent germicidal disposable wipes disinfecting towelettes or antimicrobial spray 3 Let air dry or dry with a separate clean soft cloth Notice Do not use cleaning ...

Page 13: ...ordance with hospital practices for cleaning OR equipment After every use wipe the warming unit the outside of the warming unit hose and any other surfaces that may have been touched Use a damp soft cloth and a hospital approved mild detergent germicidal disposable wipes disinfecting towelettes or antimicrobial spray 3 Let air dry or dry with a separate clean soft cloth Notice Do not use cleaning ...

Page 14: ...spenser Shipping label Marking pen METHOD 1 Call 3M technical support see page 12 for contact information to get a Return Authorization RA number and a service carton 2 Remove the top foam piece If the plastic liner is available use it to enclose the warming unit 3 Lower the warming unit carefully into the carton 4 Replace the top foam piece Observe proper orientation Hose will extend out in the c...

Page 15: ...spenser Shipping label Marking pen METHOD 1 Call 3M technical support see page 12 for contact information to get a Return Authorization RA number and a service carton 2 Remove the top foam piece If the plastic liner is available use it to enclose the warming unit 3 Lower the warming unit carefully into the carton 4 Replace the top foam piece Observe proper orientation Hose will extend out in the c...

Page 16: ... 5 GHz d 2 3 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manuf...

Page 17: ... 5 GHz d 2 3 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manuf...

Page 18: ...ct the filter 2 Return filter or replace as necessary Note Replace every 12 months or every 500 hours PN 90047 CALIBRATE THE OPERATING TEMPERATURE 1 Attach the Model 22110 temperature test unit to the end of the warming unit hose 2 Run in Ambient mode and high fan setting for 10 minutes 3 Enter the calibration mode a Press the Alternate Mode button and Ambient button together for approximately 3 s...

Page 19: ...ct the filter 2 Return filter or replace as necessary Note Replace every 12 months or every 500 hours PN 90047 CALIBRATE THE OPERATING TEMPERATURE 1 Attach the Model 22110 temperature test unit to the end of the warming unit hose 2 Run in Ambient mode and high fan setting for 10 minutes 3 Enter the calibration mode a Press the Alternate Mode button and Ambient button together for approximately 3 s...

Page 20: ...irculation d air 24 Mode Veille 24 Pannes 24 Surchauffe 25 Autres modes 25 Accès aux autres modes 25 Sortie des autres modes 25 Entretien préventif nécessaire 25 Étalonnage de la température de fonctionnement 25 Test du système de détection de surchauffe 26 Affichage du journal des codes de panne 27 Effacement du journal des codes de panne 27 Tableau des codes de panne 27 Affichage du compteur d h...

Reviews: