background image

19

78-8136-3182-3

4.  Depress the on/off button to activate the 3M

 Charge Analyzer 711.  The display will briefly indicate 

RECEIVED DATA WAIT

.

5.  The display will next indicate 

EXTERN EFM, A<OFF & ON>B

.  Depress the “B” button (ON) to activate 

the 3M

 Remote Field Sensor Probe.  Selecting the “A” button (OFF) will deactivate the sensor and return 

the Charge Analyzer to an internal measurement mode.

6.  When ON is selected the display will indicate 

EXT. EFM, D = CM, A<CHANGE, & OK>B

.  Select the 

desired measurement distance by depressing the “A” button (CHANGE) momentarily until desired distance is 

displayed.

Note:   This value represents the distance in centimeters between a charged surface and the sensor head.

7.  Depress the “B” button (OK) to select.  The display will briefly indicate 

PARAMETER CHANGED

 when 

distance value is changed.

8.  The display now indicates external field mode, distance, and measured voltage.  To change the distance 

setting again while in the external field mode, depress the “B” button momentarily and repeat steps 5-7.

9.  With red cover on probe, zero the remote field sensor by inserting a small probe or tool into the hole located at 

the rear of the sensor.  At the bottom of the hole there is a membrane switch.  Momentarily apply light pressure 

to activate the zeroing feature.  The analyzer display will briefly indicate 

ZERO ADJUSTMENT

.

Note:  When changing to a lower distance setting, re-zeroing is recommended.  For higher distances it is not 

necessary to re-zero.

This completes the remote field sensor probe setup & zeroing. 

Use

1.  Remove the red cover from the remote field sensor probe.
2.  Position the sensor in front of the charged surface or object to be measured.

Important Note

•   

When measuring an unknown voltage on a surface, position the probe at a far distance away and approach 

the surface slowly to prevent accidental discharges to the sensor element.  If there is a suspected high 

charge or if the surface of the object is uneven, the measuring distance should be increased.

3.  Observe the measured voltage (U = Volts or Kilovolts) on the analyzer display.

Note:  The remote field sensor probe measures the electrical field in V/m.  The electrical field is measured 

automatically based on the set measurement distance and is then internally calculated and indicated 

as volts in the displayed U = V.
Example 1: If the distance between the object and sensor equals 2.5 cm and the “U” value indicates 

500 volts, then the strength of the electric field is in this case 500V/inch (Since 1 inch = 2.5 cm 

approximately).

 

To change the measurement into V/m, multiply the indicated U = V value displayed by the appropriate 

factor to obtain 100 cm, (1 m) length. 

 

Example 2:  If the distance between the object and the sensor equals 10 cm and the “U” value 

indicates 500 volts, using chart below, then the V/m is in this case 5000 volts.

 

Distance (cm)

Factor

1

100

2,5

40

5

20

10

10

20

5

To convert readings into volts/inch or volts/cm use the following two formulas:

R/100 x D (cm) or R/39.4 x D (in.) where:

R = Reading indicated on the analyzer display.  D = Distance between object and sensor electrode, measured in 

inches or centimeters.

The charge analyzer 711 is now ready to perform measurements through the remote field sensor probe.

Summary of Contents for Charge Analyzer 711

Page 1: ...e Analyzer 711 Operating Instructions August 2005 78 8136 3182 3 Distributed by All Spec Industries 5228 US Hwy 421 N Wilmington NC 28401 sales all spec com www all spec com PH 800 537 0351 FX 800 379...

Page 2: ...6 3 Specifications 7 4 Fieldmeter Operation 8 5 Voltmeter Operation 10 6 Static Decay Time Balance Operation 13 7 Remote Field Sensor Probe Operation 18 8 Data Recording Analog Outputs 20 9 Verificat...

Page 3: ...ry and or property damage WARNING Indicates a potentially hazardous situation which if not avoided could result in death or serious injury and or property damage CAUTION Indicates a potentially hazard...

Page 4: ...at contains lead in the solder At the end of service life dispose of the charge analyzer in accordance with federal state and local requirements CAUTION To reduce the risks associated with property da...

Page 5: ...of the analyzer can be maintained through the AC adaptor Recharge time is approximately 14 hours with unit off when batteries are fully discharged Important Note Avoid extreme discharge of the rechar...

Page 6: ...red cover F Teflon measuring lead 1m G RS 232 interface cable H 50 mm 2 inches high conductive con tainer with insulating handle I Insulated bulldog clip J Metal spacers 3 with thread 76 mm 3 in leng...

Page 7: ...ment Ranges Distance Voltage Resolution 1 cm 0 to 10kv 1V 2 5 cm 0 to 20kv 2V 5 cm 0 to 50kv 10V 10 cm 0 to 100kv 10V 20 cm 0 to 200kv 20V Displays Two 11 segment positive negative LED bar charge indi...

Page 8: ...amaged Do not attempt to modify or repair no user serviceable parts inside contact 3M Service for repair To reduce the risks associated with fire or explosion Do not operate the analyzer in an explosi...

Page 9: ...r The digital display reads the field strength in kV m To convert readings of kV m into volts use the following formulas R 100 x D cm or R 39 37 x D in where R Reading indicated on the display and D D...

Page 10: ...zer 711 in Voltmeter mode upper left corner R Voltage measurement range selected If range selected is too low OVERFLOW will appear AR Auto voltage measurement range selected arrow symbol will briefly...

Page 11: ...l of the unit to an earth electrical ground 6 Depress the on off button to activate the analyzer The display will briefly indicate CHANGE MODE OF OPERATION then VOLTMETER IS ENABLED if cup has not bee...

Page 12: ...on a person Demonstration 3 Performance of an ESD Control Flooring 1 Stand on a non static control flooring e g mat tile epoxy etc 2 Hold the cylinder electrode in one hand and perform normal body mo...

Page 13: ...ESD control surface Demonstration 5 Discharge Performance of a Static Control Packaging Material 1 Place a packaging material on top of a grounded static control worksurface 2 Charge up the conductive...

Page 14: ...ge selection to perform the zeroing This completes the zeroing of all ranges Setting the START STOP voltage Note The Start voltage can only be adjusted from 600 to 1100 volts The Stop voltage can only...

Page 15: ...ime period The unit will not have an active timer in this mode Elapsed time must be recorded by the operator using a stop watch or other time recording device Connection to the analog output using a c...

Page 16: ...ohol Use clean cotton or latex gloves when directly handling these spacers 3 Position the ionizing air gun above the charge plate and activate the air gun blowing ionized air at the plate The timer in...

Page 17: ...tatic decay time measurement and storing of the test data is not required depress the B button The display will briefly indicate BREAK and then return to the DECAY TIME START menu Demonstration 7 Char...

Page 18: ...amaged Do not attempt to modify or repair no user serviceable parts inside contact 3M Service for repair To reduce the risks associated with fire or explosion Do not operate the analyzer in an explosi...

Page 19: ...probe setup zeroing Use 1 Remove the red cover from the remote field sensor probe 2 Position the sensor in front of the charged surface or object to be measured Important Note When measuring an unkno...

Page 20: ...der requirements Internal resistance Rin greater than 1000 ohms Connect recorder to the red and black banana jacks at the rear of the analyzer using banana plug style wires PC Requirements for Data Re...

Page 21: ...e cup electrode a 4 Connect remaining two wires between the digital voltmeter voltage inputs and the red black banana jacks analog output at the rear of the analyzer Select meter scale 2vdc 5 Select t...

Page 22: ...____________ Range Setting P S Voltage Display Voltage Pass Fail Analog Output Pass Fail 25 VDC 25Vdc Full Range 2 5 24 4 25 6 Full Range 2 0Vdc 10 100VDC 100Vdc Full Range 2 5 97 5 102 5 Full Range 2...

Page 23: ...field sensors should be covered at all times with the supplied red covers Extreme care should be taken to prevent contact to the sensor elements They should be kept free of insulating coatings caused...

Page 24: ...rized of cer of 3M Warranty Limited Remedy Limited Liability This product will be free from defects in material and manufacture for a period of one 1 year from the time of purchase 3M MAKES NO OTHER W...

Reviews: