Self Calibration
Use this function to calibrate the 945DSP anytime the
outside temperature changes by more than 35°F (20°C).
Calibrate the 945DSP at the same temperature at which it
will be used.
Short the red, green, and black leads together then press
.
You will see the following screen while the calibration is in
progress.
After a few seconds delay, the 945DSP will display the
results of the self calibration.
This symbol indicates that the self calibration was
successful. After a few moments, the unit returns
to the power on screen.
Leads open, short Red-Black-Green and
try again.
This symbol indicates that the equipment is in need
of repair. The number in the stop sign indicates to
the service department the nature of the problem.