Rinse the wand tip with clean tap water for at least one minute. Do
not
submerge the
wand.
5.
Dry the wand with a clean, lint-free wipe.
6.
Prepare the Birex solution according to the manufacturer’s recommendations.
7.
Spray and wipe the tip of the wand, from the lens to the end of the tip protector (the
silver metal section). Wet all surfaces of the tip protector completely with the Birex
solution.
8.
Allow the wand to stand for ten minutes. (The tip protector and the lens should be
completely wet by the Birex solution.)
9.
Completely wipe down the lens area with a lint-free wipe soaked in Birex.
10.
After disinfection, rinse the lens area of the wand with one liter of sterile water.
11.
Dry and barrier the wand.
Tip protector
Lens area