MIRAGE Line Compressors Instruction Manual
The compressor should be installed in a clean environment.
We recommend creating the compressed air circuit with a classic “ring” type circuit
using Rilsan Pa12 tubing (or an equivalent type of plastic compressed air tubing);
specifi cally:
Main line: Ext. D 18 mm
Secondary line / connection between ring and dental chair: Ext. D 14 mm
Electrical connection
The main connections must be made by specialized technicians. To guarantee correct
installation, check the current and voltage value of the building electrical supply. Connecting
the compressor to an electrical supply with the incorrect voltage can cause serious damage.
CAUTION
Do not replace the electrical power cable if not expressly authorised by the supplier
or by an authorised installer: replacing the cable could lead to interference with other
nearby equipment and/or faults.
Beware of the risk of electric shock!
Disconnect the power at the main disconnect switch before connecting the compressor.
Install the equipment in a place where it cannot come into contact with water or other
liquids.
Make sure the electrical power line is suffi cient for the compressor, following the data shown
on the motor rating plate.
Failure to comply with these instructions could lead to death, fi re, or
electrocution.