23
discharged in waste line and air will be expelled outside of the building through the
exhausting pipe.
•
Class I equipment
•
Ordinary equipment (IPX0)
•
Equipment not suitable for use in the presence of a flammable anesthetic mixture
with air or with oxygen or nitrous oxide
•
Continuous operation
This equipment has been tested and found to comply with the limits for medical devices in
IEC 60601-1-2. These limits are designed to provide reasonable protection against harmful
interference in a typical medical installation. This equipment generates, uses and can radiate
radio frequency energy and, if not installed and used in accordance with the instructions,
may cause harmful interference to other devices in the vicinity. However, there is no
guarantee that interference will not occur in a particular installation.
If this equipment does cause harmful interference to other devices, which can be determined
by turning the equipment off and on, the user is encourage to try to correct the interference
by one or more of the following measures:
•
Reorient or relocate the receiving device
•
Increase the separation between the equipment
•
Connect the equipment into an outlet on a circuit different from that to which the other
device(s) are connected
•
Consult that manufacturer or field service technician for help
3.
Installation notes
3.1.
Environmental conditions
The pump is rated for indoor use only.
Maximum altitude 2000 meters.
Operating temperature range 10°C to 40°C.
Maximum relative humidity of 80% for temperature up to 31°C decreasing to 50% at 40°C.
Rated for +/- 10% of supply voltage.
Pollution degree 2.
Installation category II
3.2.
Introduction
This manual has been compiled not only for the care and maintenance of the suction system
now in your possession, but as a helpful reference and guide to prevent many problems
which can occur if used improperly.
3.3.
Transport & storage conditions
Temperature: -20°C to 40°C
Humidity: 0% to 90%
Pressure: 700hPa to 1060hPa
3.4.
Unpacking
Carefully remove the machine from the shipping case. Preserve all paperwork for future
reference. If damage has occurred from shipment a claim must be filed with the carrier
immediately; preserve the shipping carton for inspection by the carrier. If you are required
to communicate with your dealer, be sure to include serial numbers for quick identification.
Summary of Contents for RAIN2+
Page 35: ...35 ...
Page 36: ...36 M R12 rev 02 4th May 2019 ...