4.2  Set-up, Stand-by, Power-down, Buzzer 

 

To enter into set-up mode for stand-by, power-down and audible warning, press 
the M  key for 10 seconds, then release. 

 

The  display  shows  the  symbol  “L”  (flashing)  and  the  last  set  stand  by  time 
(displayed as minutes). 

The  stand-by  time  is  the  time  after  which    the  heater  temperature  is 
automatically reduced to 60°C when the unit is left idle. 

The stand-by time can be decreased or increased with the 

 

 keys  

 

“no” stand-by or between 1 minute and 99 minutes with steps of 1minute. 

 

By  pressing  the  M    key  again  the  new  stand  by  time is stored  and  the  power-
down time can be set next. 

The  display  shows  the  symbol  “P”  (flashing)  and  the  last  set  power-down  time 
displayed as hours). 

The  power-down  time  is  the  time  after  which  the  heater  is  switched  off 
automatically when the unit is left idle. 

 

The power down-time can be decreased or increased with the 

 keys  

 

between 1 hour and 24 hours with steps of 1 hour. 

 

By pressing the M  key again the new power down time value is stored and the 
audible warning (buzzer) can be set next. 

The display shows the symbol “S” (flashing) and the last set buzzer function. 

The  buzzer  function  can  be  enabled  (“on”)  or  disabled  (“no”)  with  the 

 

 

keys. 

 

By pressing the  M   key again the buzzer function is stored and the unit returns 
into  “run  mode”  (see  page  12)  and  the  actual  temperature  of  the  heating 
element is shown in the display. 

 

 

     11 

Summary of Contents for 4s2

Page 1: ...4s2 Variable Temperature Thermal Sealer foto...

Page 2: ...references to names or parts are purely casual and are used to illustrate the product described The document is for information only The manufacturer accepts no liability for errors contained herein...

Page 3: ...Operating Mode 10 4 2 Set Up Stand By Power Down Buzzer 11 5 Basic Functions 12 5 1 Run Mode 12 5 2 Stand By Mode 14 5 3 Power Down Mode 14 5 4 Drawer Locking 14 6 Troubleshooting 15 6 1 Introduction...

Page 4: ...the operator needs to position the sample plate together with a foil or film on the plate carrier The sealing operation starts once the drawer with the plate carrier has been closed Film and sample pl...

Page 5: ...wing safety notes This product must only be used in accordance with proper safety standards and procedures together with the instructions included in this manual The sealer has been designed and manuf...

Page 6: ...the unit is only connected to an earthed supply There are no user accessible or serviceable parts inside the unit Do not remove the unit s cabinet WARNING Hot surface The hot plate can reach temperat...

Page 7: ...ve humidity 20 80 not condensing SET UP PARAMETERS Sealing temperature OFF or from 120 C to 200 C Step 1 C Default OFF Sealing time up to 10 seconds Step 0 5 seconds Default 3 seconds Stand by cycle n...

Page 8: ...FICATION 2 1 Sealer Parts Casing 1 Control panel 2 Rear Power Interface 3 Heating element 4 Plate carrier 5 96well PCR Plate Adapter 6 optional Weighted platten 7 Sealing Frame 8 Chapter 2 5 2 1 3 4 5...

Page 9: ...Control panel Figure 2 Control Panel Display Three Digit LED display Keys 1 M Set up Run mode key 2 Directional key up 3 Directional key down Led 4 Status Indicator Amber 1 2 3 4 6...

Page 10: ...2 1 2 Rear Power Interface Figure 3 Rear Power Interface 1 Power Setting 115 230 Vac and Fuse Holder 2 I O Power Switch 3 Mains Power Plug 2 1 3 7...

Page 11: ...ltage selector switch should be checked to ensure that the voltage has been set at the correct value Connect the unit to the mains power supply and switch on the switch I O of the rear power interface...

Page 12: ...rewdriver lever out the fuse holder figure 3B 3 Replace the two fuses according to the table below Power Voltage EN 60127 115v 3 15A T 230v 2A T Table 4 Insert the fuse holder keeping the set voltage...

Page 13: ...see page 20 for recommended sealing temperatures By pressing the M key again the new temperature value is stored and the sealing time can be set next The display shows the symbol t and the last set se...

Page 14: ...n time can be set next The display shows the symbol P flashing and the last set power down time displayed as hours The power down time is the time after which the heater is switched off automatically...

Page 15: ...n be loaded onto the standard plate carrier However for easier handling of shallow well plates we recommend to place the standard plate carrier 4ti 0615 inside the adapter for 96well PCR plates option...

Page 16: ...stop moving An error message will be displayed and a continuous audible beep will be emitted To rectify this and to continue with the sealing process push the drawer in fully until you can a beep If...

Page 17: ...e stand by function 5 3 Power down Mode If the unit remains idle in stand by mode for a prolonged time the system will go into power down mode and the heating element will be switched off factory set...

Page 18: ...the low position within a defined time Er5 motor over current Er6 temperature sensor opened or disconnected Er7 open drawer sensor active during elevation movement Er8 platform in low position sensor...

Page 19: ...supply Do not attempt to clean the unit until it has cooled down to near room temperature Release the lever on the back of the unit counterclockwise see figure The drawer can now be fully removed fro...

Page 20: ...Reinsert the drawer into the unit Move the lever on the back of the unit clockwise to secure the drawer mechanism see figure Switch ON the unit 17 Figure 8 C Lever in open position...

Page 21: ...should not be immersed in solvents Do not use acetone or abrasive cleaners No parts are to be autoclaved In case of radioactive spillages it is recommended to use an appropriate cleaning agent For any...

Page 22: ...ene laminate Clear polyester polypropylene laminate Clear polymer Thin clear polymer Foil laminate Foil 8 3 Plate Carriers Accessories Standard Plate Carrier for skirted plates included product code 4...

Page 23: ...nd temperature for a 348 Well Plate 4ti 0521 Peel Seal 175 C 3sec 175 C 2sec 4ti 0531 Pierce Seal 175 C 2 5sec 175 C 2sec 4ti 0533 Pierce Seal Strong 175 C 2 5sec 175 C 2sec 4ti 0536 Foil Seal 175 C 2...

Page 24: ...the sealer 2 Switch off the unit 3 Pull out the drawer 4 Place card board sheet provided between the underside of the drawer and the top of the lifting platform inside the unit 5 Push drawer in Note T...

Page 25: ...ves note of defects or non conformance during the warranty period the manufacturer will at its option repair or replace the affected product The manufacturer will provide at no additional charge parts...

Reviews: