User Manual
LIOSMART 100
28
1-4: Temperature shelve probe correction
4 corrections points are defined. (-40, 0,+30,+70)
For each correction point correction can be set. To change value use .
5-8: Material temperature probe correction
For material temperature
probe
, 4 corrections points are defined.
For each correction point correction can be set. To change value use .
9: Vacuum display
Pc … Current chamber pressure
Pp … Current vacuum pump pressure
10: Time count-down tolerance (temperature)
You can set the time of each temperature step. The countdown will start when
temperature is in set limits.
For example:
Set temperature: 50°C
Set time: 60 min
Set START TIME COUNT IN RANGE : +- 1°C
The countdown will start when the temperature will be between 49.0°C and 51.0°C !
11-15: Information valid only for service personal!
16: Communication port setup
17: Selection of menu language
English
Slovensko
Deutsch
18: Display contrast setup