background image

- 6 -

- 7 -

Chapter 8  Control Test

This meter checks itself every time you insert a test strip. You do not need to 

run frequent control tests. Do a control test if:

n

 You suspect the meter or test strips are not working properly.

n

 You dropped the meter.

n

 You damaged the meter.

n

 Your test results do not match how you feel.

n

 You want to check performance of the meter and test strips before a blood 

glucose test. 

NOTE

 

Use only the control solution, GLUCOCARD

TM

 MX CONTROL, to test your 

meter and test strips.

 CAUTION

n

 

Do not

 use control solutions beyond their expiration date. The expiration 

date is written on the control solution bottles next to “Exp.”. 

n

 Leave the meter and test strips and the control solution at 

   10 to 40°C (50 to 104°F) and 20 to 80% RH 

  for at least 30 minutes before running the control test. 

n

 

Do not

 drink control solution. It is not for human consumption. 

n

 Keep control solution away from skin and eyes. Contact may cause 

inflammation.

1. See steps 1 and 2 in chapter 2.7.

2. Select the control test flag.  

As an alternative, you can attach 

the control test flag after the test result appears on the screen. Test results with 

the control test flag are not part of your result averages.

3. Make sure that the blood symbol is flashing on the screen. 

4. Squeeze a drop of control solution to the tip of the test strip.

The test will not start if you apply your control solution to 

the check window directly.

The test starts when the meter detects the control 

solution. The meter counts down from 5 to 1 during the 

test. 

NOTE

 

n

 Tightly close the control solution bottle.

n

 

Do not

 touch the test strip once the meter starts counting down. 

5. Check your test result is within the acceptable range. 

The acceptable range is shown on the label of test strip bottles. If within range, 

the meter and the test strip are working properly. If out of range, test the control 

solution again.

 CAUTION

If you still have test results that are out of the acceptable range, stop using 

your meter, test strips and control solutions, and contact your distributor.

6. Slide the test strip disposal lever to eject the test strip. 

The meter turns itself off.

Hold for 

2 seconds

Chapter 9  Meter Care

9.1 Storing Your Meter

After use, tightly close the caps of the test strip bottle and control solution bottle 

to maintain their quality. 

Store your meter, test strips, control solutions and manuals in your carrying 

case and store it in a dry location. The correct storage temperature is 0 to 50°C 

(32 to 122°F) for the meter and 1 to 30°C (34 to 86°F) for the test strips and 

control solutions. 

Do not

 freeze. Avoid heat, humidity and direct sunlight.

 CAUTION

To obtain accurate test results: 

n

 

Do not

 use test strips or control solutions if their bottles are broken or 

have been left open. 

n

 

Do not

 use the test strips or control solutions beyond their expiration 

dates.

9.2 Cleaning Your Meter and Lancing Device

Your meter and lancing device do not need special cleaning. If your meter and 

lancing device get dirty, wipe them with a soft piece of cloth moistened with 

water. To disinfect these items after cleaning, wipe with a soft cloth moistened 

with 70% ethanol, 70% isopropanol or diluted household bleach (0.05% sodium 

hypochlorite solution).

9.3 Changing the Batteries  

When the battery symbol appears on the screen, the batteries 

are getting low. Before using your meter, change the batteries. 

Past results remain in the memory even when the batteries are 

changed.

NOTE

 

Your meter uses two CR2032 3V lithium batteries. This type of batteries is 

available in many stores. Keep spare batteries handy at all times.

You do not need to set the date and time if you insert new batteries within 1 

minute after removing the old batteries. Even if you change the batteries within 

1 minute, your meter’s clock may stop if:

n

 The batteries are not replaced soon after the battery symbol appears.

n

 The batteries are inserted upside down.

n

 You touch the metal parts inside the meter with hands or metal.

1.

 Make sure the meter is off. 

2.

 Grip the strap hole with the tip of your finger and pull 

open. 

3.

 Remove the batteries.

4.

 Set the new batteries into the battery holder with the 

“+” side facing upward.

NOTE 

If the batteries are inserted upside down, the meter will 

not operate.

5.

 Close the battery holder. 

6.

 The “SET” screen appears when the time has been reset. If it appears, set 

the correct time. See the setup mode flowchart in chapter 7. If you do not set 

the time, all subsequent test results will be stored as 12:00 AM or 00:00.

Dispose of old batteries according to your local environmental 

regulations.  

Summary of Contents for GLUCOCARD MX

Page 1: ...induce severe low glucose symptoms such as coma 2 2 What You Need for Each Test a GLUCOCARDTM MX meter b GLUCOCARDTM MX BLOOD GLUCOSE TEST STRIPS c Lancing device d Lancet WARNING Keep the meter test...

Page 2: ...test strip into the meter Double beeps begin to sound 4 minutes and 30 seconds after inserting the test strip Thirty seconds later the meter turns itself off NOTE In the ECO mode the blood symbol flas...

Page 3: ...d The meter counts down from 5 to 1 during the test NOTE n For accurate test results touch the test strip to your blood in less than 20 seconds after puncturing n Do not test blood that runs or spread...

Page 4: ...yond the 500th overwrite previously stored results in the order of oldest first 1 Make sure the meter is off 2 Turn on the meter The most recent test result appears on the screen The 3 digit number fo...

Page 5: ...ucose When the clock counter setting is on the elapsed time and after meal flag appear on the right of the test result 2 To change or set the elapsed time Select the time Confirm your setting 3 Slide...

Page 6: ...old To make another alarm setting repeat steps 1 to 4 To the top of the beeper volume setting To the top of the date setting The meter turns off NOTE The meter turns itself off if left alone for 3 min...

Page 7: ...toring Your Meter After use tightly close the caps of the test strip bottle and control solution bottle to maintain their quality Store your meter test strips control solutions and manuals in your car...

Page 8: ...accuracy Within 75 seconds month Operating Temperature 10 to 40 C 50 to 104 F environment Humidity 20 to 80 RH No condensation Expected life 3 years according to company data Dimensions L102 W33 18 5...

Page 9: ...Operating Manual MASTER 43125 12 11 420 x 300 mm BLACK K 100...

Reviews: