Peel from this end
50
Remo
ving a sensor
/
English
3.
Pull back on the transmitter release
tab.
4.
Lift the transmitter away from the
sensor adhesive patch.
5.
Place the transmitter in the charger.
Make sure the transmitter is fully
charged before using it again.
See SECTION 8.4 for details about
charging the transmitter.
6.
Discard the used sensor adhesive
patch as medical waste.
CAUTION
Do not discard your transmitter.
Once the session has stopped, you should remove the sen-
sor and place the transmitter in the charger.
See CHAPTER 13 for details about removing your sensor.
/ 13
Removing a sensor
Once a session stops, you should remove the sensor and
place the transmitter in the charger.
CAUTION
•
The sensor is for single-use only, up to 14 days. Do not
reuse the sensor or use it beyond 14 days.
• Do not remove the transmitter while the sensor assem-
bly is attached to you.
1.
Remove the sensor adhesive patch, starting at the end closest to
the flat back side of the transmitter.
Peel the patch away from your body.
2.
After removal, fold back the sides of
the sensor adhesive patch to cover
the sensor.
Summary of Contents for GlucoMen Day CGM
Page 1: ...User Guide ...
Page 67: ......