59
Additional Informa
tion
/
English
• Be careful and gentle while cleaning the transmitter con-
tacts.
17.3
System disposal
Consult your local regulations when disposing of the sen-
sor insertion tool, the sensor guide, the transmitter, and
the charger.
Dispose of the sensor adhesive patch as medical waste.
17.4
Product specifications
17.4.1
Sensor specifications
Specification
Description
Glucose range
2.2–22.2 mmol/L
Sensor life
Up to 14 days
Calibration
GlucoMen Day METER (provided
separately) or commercially avai-
lable blood glucose meter.
Calibration range
3.9–13.9 mmol/L
Storage temperature
5°C to 35°C (41°F to 95°F)
Storage humidity
10% - 90% Relative humidity
Operating temperature
5°C to 45°C (41°F to 113°F)
Sterilization
Sterile by irradiation
Altitude
Up to 3048 metres (10,000 feet)
Atmospheric pressure
700 hPa - 1060 hPa
17.4.2
Transmitter specifications
Specification
Description
Transmitter battery type
Non-serviceable, rechargeable lithium-ion
polymer battery
Transmitter battery
life
14 days of typical use
Transmitter life
5 years of average continuous use
Transmitter data
storage
14 days of calculated glucose data,
sampled once per minute
Transmitter dimensions
Length: 33.0 mm, width: 24.0 mm,
thickness: 9.0 mm
Transmitter weight
5 grams
Ingress Protection
IP27: Protection against insertion of large
objects and immersion in 1 metre (3 feet) of
water for up to 30 minutes.
Operating temperature
5°C to 45°C (41°F to 113°F)
Operating humidity
10% - 100% Relative humidity, condensing
Storage temperature
5°C to 45°C (41°F to 113°F)
Storage humidity
10% - 100% Relative humidity, condensing
Protection against
electrical shock
Type BF applied part
TX Frequency
2.402–2.480 GHz
Reading Frequency
Every 1 minute
Bandwidth
1.06 MHz
Maximum Output Power
-2.93 dBm
Modulation
Gaussian FSK
Altitude
Up to 3048 metres (10,000 feet)
Atmospheric pressure
700 hPa - 1060 hPa