Pulse Output
Sets and controls action of the
programmable Pulse Output
cir
cuitr
y.
NOTE:
Unit/Pulse value has to be entered
in currently selected EU.
EU has to be not time based
It is recommended to set the unit/
pulse value equal the device
maximum fl
ow in the same units per
second equivalent. This will limit the
pulse to no faster than one pulse ever
y
second.
Example:
Maximum fl
ow rate:
600 liter/min (600 liter/min = 10 liters
per second)
If Unit/Pulse is set to 10 liters per
pulse, the output will pulse once ever
y
second (F=1 Hz).
Pulse active time in ms has to be at
least twice less than pulse period (1/F).
In this example any value between 50
and 500 ms will be acceptable.
18
P
U
Set Units Per Pulse
Parameter
.
<V
alue>
(Unit/Pulse)
In current E.U.
(example:
10 litr/pulse)
PU:<value>
Example:
PU:10
T
Set Pulse active
T
ime in ms
<value>
[25-3276 ms]
PT
:<value>
Example: PT
:100
D
(disable pulse
output)*
P:D
E
(enable pulse
output)
P:E
Q
(read current pulse
output Queue value)
PQ:<value>
(number of pulses in
Queue)
F
Set Flow Start value
<value>
(0.0–100.0%FS)
PF:1.0
S
(read setting status)
PS:Mode,FlowStart,
Unit/ Pulse,Pulse T
ime Inter
val
Example:
PS:E,1.0,1.666,100