50
e) Totalizer Auto Reset (Tabular entry)
This option allows the automatic reset of the Totalizer when it reaches a preset Action
Volume value. This feature may be convenient for batch processing, when a predefi ned
volume of fl uid must be repeatedly dispensed into the process. Two selections are
available: Enabled or Disabled. The default entry is Disabled. Totalizer Auto Reset
selections can be set with the joystick
UP
and
DN
buttons and are accepted by
pressing the joystick equivalent of an
ENT
button.
f) Totalizer Auto Reset Delay (Numerical entry)
This option may be desirable when the
“Totalizer Auto Reset”
feature is enabled and
a predefi ned delay is required before a new batch cycle starts. Valid settings are in the
range of 0 to 3600 seconds. The default value is 0: no delay.
g) Reset Totalizer (Tabular entry)
Either Totalizer’s reading can be reset by selecting the
“Reset Totalizer”
menu option.
A typical Totalizer Reset screen is shown below:
Figure 31:
Reset Totalizer Screen
When
the
“YES”
option is selected, Totalizer #1 will be reset. A confi rmation screen
will appear (see
Figure 32
).
**************************
Totalizer
has been reset
**************************
Press any key...
Figure 32:
Totalizer Reset Confi rmation
Reset Totalizer #1:
NO
YES
DO YOU WANT
RESET TOTALIZER?
Summary of Contents for DPC
Page 6: ...2...
Page 120: ...116 APPENDIX I COMPONENT DIAGRAM Top Component Side...
Page 121: ...117 Bottom Component Side...